HIGH-THROUGHPUT TITRATION VIA SEGMENTED FLOW IN A LAB-ON-A-CHIP DEVICE FOR ENVIRONMENTAL SENSING APPLICATIONS
We introduce a high-throughput titration technique within a lab-on-a-chip (LOC) system, reducing sample usage and assay time, and demonstrate its application for pH determination necessary in total alkalinity (TA) measurements.LOC devices, including an autonomous opto-fluidic analyzer developed by our group [1], enabled conducting in-situ titration-based assays, crucial for water quality monitoring through measuring various parameters, e.g., TA and dissolved inorganic carbon (DIC) [2,3].In LOC-based titration, each point usually involves filling the entire device with a sample and titrant, maintaining a single dilution factor, = + , and
Authors
- Vincent J. Sieben (ORCID: https://orcid.org/0000-0002-6289-9440)
- Shahrooz Motahari
- Andre Hendricks (ORCID: https://orcid.org/0000-0003-2337-7284)
- Colin Sonnichsen (ORCID: https://orcid.org/0000-0003-4769-1681)
- Alireza Zabihihesari
Institutions
- Dalhousie University (CA)
Publication Details
- Published
- 2024-10-07
- DOI
- https://doi.org/10.70477/lxzj4627
- Primary Topic
- Microfluidic and Bio-sensing Technologies
- Type
- article
- Field-Weighted Citation Impact
- 0.00