Performance of a focused 2D anti-scatter grid for industrial X-ray computed tomography

X-ray computed tomography (XCT) is increasingly being used for the measurement and inspection of large dense metallic engineering components. When scanning such components, the quality of the data is degraded by the presence of scattered radiation. In this work, the performance of a focused 2D anti-scatter grid (ASG) is investigated for scanning samples made from cobalt chrome and Inconel on a 450 kV cone-beam XCT system. The devised scatter correction method requires one additional scan of the sample, and for projections to be algorithmically processed prior to reconstruction. The results show that the ASG based scatter correction method increases the contrast-to-noise of the data by 14.5% and 61.5% for the cobalt chrome and Inconel samples, respectively. Furthermore, the method increases edge sharpness by 6% and 16.9% for outer and inner edges, respectively.

Authors

Institutions

Publication Details

Journal
Journal of X-Ray Science and Technology
Published
2025-03-25
DOI
https://doi.org/10.1177/08953996251325072
Primary Topic
Advanced X-ray and CT Imaging
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Performance of a focused 2D anti-scatter grid for industrial X-ray computed tomography

Joseph Lifton, Zheng Jie Tan, Christian Filemon
Journal of X-Ray Science and Technology
Advanced X-ray and CT Imaging
article

Performance of a focused 2D anti-scatter grid for industrial X-ray computed tomography

Joseph Lifton, Zheng Jie Tan, Christian Filemon
article en

Abstract

X-ray computed tomography (XCT) is increasingly being used for the measurement and inspection of large dense metallic engineering components. When scanning such components, the quality of the data is degraded by the presence of scattered radiation. In this work, the performance of a focused 2D anti-scatter grid (ASG) is investigated for scanning samples made from cobalt chrome and Inconel on a 450 kV cone-beam XCT system. The devised scatter correction method requires one additional scan of the sample, and for projections to be algorithmically processed prior to reconstruction. The results show that the ASG based scatter correction method increases the contrast-to-noise of the data by 14.5% and 61.5% for the cobalt chrome and Inconel samples, respectively. Furthermore, the method increases edge sharpness by 6% and 16.9% for outer and inner edges, respectively.

Journal of X-Ray Science and TechnologyVol. 33(3)
Agency for Science, Technology and Research (SG), Nanyang Technological University (SG), University of Southampton (GB)
Agency for Science, Technology and Research
Industry, innovation and infrastructure
Openalex Percentile: Top 95%
Advanced X-ray and CT Imaging
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.