Investigation of Radiation Isocenter Accuracy of Linear Accelerator – New Approach of Comprehensive Evaluation of Gantry, Couch, and Collimator Isocenters
This combined radiation isocenter verification tool aids in comprehensive evaluation of gantry, couch and collimator isocenters using radiographic film. It is easier to implement and faster to analyse compared to existing techniques, using a PC-based program that minimizes human error and increases measurement accuracy. This new approach can be used for routine quality assurance checks.
Authors
- Swathi Srinidhi (ORCID: https://orcid.org/0009-0006-7700-9486)
- Tamilarasan Mani
- P. Manikandan (ORCID: https://orcid.org/0000-0002-6513-9690)
- G. V. Subrahmanyam
- Avinash Kuppa Rao (ORCID: https://orcid.org/0009-0003-8778-0045)
Institutions
- Panacea Biotec (India) (IN)
Publication Details
- Journal
- Journal of Medical Physics
- Published
- 2025-01-01
- DOI
- https://doi.org/10.4103/jmp.jmp_131_24
- Primary Topic
- Advanced Radiotherapy Techniques
- Type
- article
- Field-Weighted Citation Impact
- 0.00