Hemolytic Interference on Blood Gas Analysis

Background: Hemolysis is an important preanalytical factor that influences laboratory test results.Because arterial blood gas analysis (ABGA) is performed using whole blood, it is difficult to visually check whether a specimen is hemolyzed, and even blood gas analyzers cannot detect hemolysis.However, there is insufficient consensus on the parameters that are influenced by hemolyzed specimens.This study aimed to determine the effect of hemolysis on ABGA results.Methods: One hundred residual arterial blood specimens were collected from Severance Hospital between March and April 2022.Samples were aliquoted into three groups for mechanical hemolysis.Hemolysis was induced using 16-, 22-, and 26-gauge needles and measured using the Profile pHOx Ultra Blood Gas Analyzer (Nova Biomedical, USA).The remaining blood was centrifuged, and the hemolysis index was determined using the plasma.Results: Among the parameters, pH and K increased, whereas pCO 2 , Na, Ca 2+ , and HCO 3 -decreased.The values of Hb, Mg 2+ , and Hct did not change with the degree of hemolysis, although there was a difference between the two groups.The values of pCO 2 , Hb, K, and Ca 2+ increased as the degree of hemolysis increased, with % biases exceeding the desirable bias.Conclusions: This study confirmed that hemolysis significantly influences pH, pCO 2 , and K. Therefore, when clinical findings and blood gas analysis results are inconsistent, clinicians should be cautious of spurious hemolysis when interpreting the results.

Authors

Institutions

Publication Details

Journal
Journal of Laboratory Medicine and Quality Assurance
Published
2025-03-24
DOI
https://doi.org/10.15263/jlmqa.2025.47.1.23
Primary Topic
Clinical Laboratory Practices and Quality Control
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Hemolytic Interference on Blood Gas Analysis

Hanmil Jang, Jong‐Baeck Lim, Sang‐Guk Lee, Hyein Kang et al.
Journal of Laboratory Medicine and Quality Assurance
Clinical Laboratory Practices and Quality Control
article

Hemolytic Interference on Blood Gas Analysis

Hanmil Jang, Jong‐Baeck Lim, Sang‐Guk Lee, Hyein Kang, John Hoon Rim
article en

Abstract

Background: Hemolysis is an important preanalytical factor that influences laboratory test results.Because arterial blood gas analysis (ABGA) is performed using whole blood, it is difficult to visually check whether a specimen is hemolyzed, and even blood gas analyzers cannot detect hemolysis.However, there is insufficient consensus on the parameters that are influenced by hemolyzed specimens.This study aimed to determine the effect of hemolysis on ABGA results.Methods: One hundred residual arterial blood specimens were collected from Severance Hospital between March and April 2022.Samples were aliquoted into three groups for mechanical hemolysis.Hemolysis was induced using 16-, 22-, and 26-gauge needles and measured using the Profile pHOx Ultra Blood Gas Analyzer (Nova Biomedical, USA).The remaining blood was centrifuged, and the hemolysis index was determined using the plasma.Results: Among the parameters, pH and K increased, whereas pCO 2 , Na, Ca 2+ , and HCO 3 -decreased.The values of Hb, Mg 2+ , and Hct did not change with the degree of hemolysis, although there was a difference between the two groups.The values of pCO 2 , Hb, K, and Ca 2+ increased as the degree of hemolysis increased, with % biases exceeding the desirable bias.Conclusions: This study confirmed that hemolysis significantly influences pH, pCO 2 , and K. Therefore, when clinical findings and blood gas analysis results are inconsistent, clinicians should be cautious of spurious hemolysis when interpreting the results.

Journal of Laboratory Medicine and Quality AssuranceVol. 47(1)
Yonsei University (KR), Keimyung University (KR)
Openalex Percentile: Top 93%
Clinical Laboratory Practices and Quality Control
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Hemolytic Interference on Blood Gas Analysis — Hanmil Jang, Jong‐Baeck Lim, et al. · Journal of Laboratory Medicine and Quality Assurance (2025) | TGRS Research Map | TGRS