Impact of Type A Standard Measurement Uncertainty in the Primary Reference Measurement Procedure for Creatinine Assay on Bias Analysis in Accuracy-Based Creatinine Proficiency Testing

Background:The accuracy-based creatinine (ABCr) proficiency testing (PT) program by the Korean Association of External Quality Assessment Service evaluates bias and acceptability in participating laboratories using target values assigned by the primary reference measurement procedure (RMP).This study evaluated the impact of type A standard measurement uncertainty (MU) in the primary RMP on bias and acceptability assessments.Methods: ABCr PT data from 2020 to 2021 were analyzed.Using the target mean, standard deviation, and number of replicates provided by the Reference Material Institute for Clinical Chemistry Standards laboratory (Japan), the type A MU for each sample was calculated using a bottom-up approach.Bias and acceptability of creatinine assays were compared with and without the inclusion of type A MU when assigning target values. Results:The type A MU of the primary RMP for the creatinine assays ranged from 0.01 to 0.02 mg/dL depending on the sample.Without type A MU, the mean bias of creatinine assay ranged from -2.9% to 6.5%.Incorporating type A MU into target values, adjusted the mean bias to -3.2% to 5.3% (with addition) and -2.6% to 7.8% (with subtraction).Acceptability rates ranged from 69.2% to 95.0% without type A MU and shifted to 70.1% to 95.1% (with addition) and 68.5% to 95.4% (with subtraction).Conclusions: The inclusion of type A MU in the primary RMP affects bias and acceptability assessment, highlighting its significance in the interpretation of ABCr PT results.

Authors

Institutions

Publication Details

Journal
Journal of Laboratory Medicine and Quality Assurance
Published
2025-03-24
DOI
https://doi.org/10.15263/jlmqa.2025.47.1.14
Primary Topic
Clinical Laboratory Practices and Quality Control
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Impact of Type A Standard Measurement Uncertainty in the Primary Reference Measurement Procedure for Creatinine Assay on Bias Analysis in Accuracy-Based Creatinine Proficiency Testing

Tae‐Dong Jeong
Journal of Laboratory Medicine and Quality Assurance
Clinical Laboratory Practices and Quality Control
article

Impact of Type A Standard Measurement Uncertainty in the Primary Reference Measurement Procedure for Creatinine Assay on Bias Analysis in Accuracy-Based Creatinine Proficiency Testing

Tae‐Dong Jeong
article en

Abstract

Background:The accuracy-based creatinine (ABCr) proficiency testing (PT) program by the Korean Association of External Quality Assessment Service evaluates bias and acceptability in participating laboratories using target values assigned by the primary reference measurement procedure (RMP).This study evaluated the impact of type A standard measurement uncertainty (MU) in the primary RMP on bias and acceptability assessments.Methods: ABCr PT data from 2020 to 2021 were analyzed.Using the target mean, standard deviation, and number of replicates provided by the Reference Material Institute for Clinical Chemistry Standards laboratory (Japan), the type A MU for each sample was calculated using a bottom-up approach.Bias and acceptability of creatinine assays were compared with and without the inclusion of type A MU when assigning target values. Results:The type A MU of the primary RMP for the creatinine assays ranged from 0.01 to 0.02 mg/dL depending on the sample.Without type A MU, the mean bias of creatinine assay ranged from -2.9% to 6.5%.Incorporating type A MU into target values, adjusted the mean bias to -3.2% to 5.3% (with addition) and -2.6% to 7.8% (with subtraction).Acceptability rates ranged from 69.2% to 95.0% without type A MU and shifted to 70.1% to 95.1% (with addition) and 68.5% to 95.4% (with subtraction).Conclusions: The inclusion of type A MU in the primary RMP affects bias and acceptability assessment, highlighting its significance in the interpretation of ABCr PT results.

Journal of Laboratory Medicine and Quality AssuranceVol. 47(1)
Ewha Womans University (KR)
Quality Education
Openalex Percentile: Top 93%
Clinical Laboratory Practices and Quality Control
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Impact of Type A Standard Measurement Uncertainty in the Primary Reference Measurement Procedure for Creatinine Assay on Bias Analysis in Accuracy-Based Creatinine Proficiency Testing — Tae‐Dong Jeong · Journal of Laboratory Medicine and Quality Assurance (2025) | TGRS Research Map | TGRS