Verification of the analytical performance of the serum Beta-2 microglobulin (β2M) assay on the Abbott Alinity ci®: Experience of the Biochemistry Laboratory of the Mohammed VI University Hospital, Oujda

The verification of analytical performance is a critical step in ensuring the reliability and accuracy of laboratory assays. This study aims to evaluate the analytical performance of the beta-2 microglobulin (β2M) immunoassay on the Abbott Alinity ci® platform in the central laboratory of Mohammed VI University Hospital of Oujda over a 30-day duration. The study assesses key analytical parameters, including precision, reproducibility, and repeatability against established standards, based on ISO 15189 accreditation guidelines. Results show low coefficients of variation (CV), with reproducibility ranging from 2.36% to 2.94% and repeatability from 1.00%to 2.55%, meeting the stringent criteria set by recognized professional societies. These findings confirm the robustness and reliability of the method and validate the analytical performance of the β2M assay, underscoring its suitability for clinical use.

Authors

Publication Details

Journal
GSC Biological and Pharmaceutical Sciences
Published
2025-03-24
DOI
https://doi.org/10.30574/gscbps.2025.30.3.0116
Primary Topic
Chronic Lymphocytic Leukemia Research
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Verification of the analytical performance of the serum Beta-2 microglobulin (β2M) assay on the Abbott Alinity ci®: Experience of the Biochemistry Laboratory of the Mohammed VI University Hospital, Oujda

Karim Jamal, Mohammed Choukri, El Houcine Sebbar, Oumaima kharkhach et al.
GSC Biological and Pharmaceutical Sciences
Chronic Lymphocytic Leukemia Research
article

Verification of the analytical performance of the serum Beta-2 microglobulin (β2M) assay on the Abbott Alinity ci®: Experience of the Biochemistry Laboratory of the Mohammed VI University Hospital, Oujda

Karim Jamal, Mohammed Choukri, El Houcine Sebbar, Oumaima kharkhach, Dounia Elmoujtahide
article en

Abstract

The verification of analytical performance is a critical step in ensuring the reliability and accuracy of laboratory assays. This study aims to evaluate the analytical performance of the beta-2 microglobulin (β2M) immunoassay on the Abbott Alinity ci® platform in the central laboratory of Mohammed VI University Hospital of Oujda over a 30-day duration. The study assesses key analytical parameters, including precision, reproducibility, and repeatability against established standards, based on ISO 15189 accreditation guidelines. Results show low coefficients of variation (CV), with reproducibility ranging from 2.36% to 2.94% and repeatability from 1.00%to 2.55%, meeting the stringent criteria set by recognized professional societies. These findings confirm the robustness and reliability of the method and validate the analytical performance of the β2M assay, underscoring its suitability for clinical use.

GSC Biological and Pharmaceutical SciencesVol. 30(3)
Openalex Percentile: Top 92%
Chronic Lymphocytic Leukemia Research
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.