Early displacement of an atrial lead: A fatal presentation of a relatively frequent complication

An 81-year-old woman was referred to Parc Taulí Hospital Universitari with complete atrioventricular block and a dual-chamber pacemaker was implanted. She subsequently presented an episode of ventricular tachycardia that degenerated into ventricular fibrillation requiring cardiopulmonary resuscitation. After review of the device, as well as the electrocardiogram and chest X-ray, displacement of the atrial lead into the right ventricle was confirmed. Contrary to expectations, the ventricular event was not secondary to endocardial irritation by the displaced lead, so the hypothesis of a possible depolarization-repolarization heterogeneity as a cause of the increased arrhythmogenicity was raised.

Authors

Institutions

Publication Details

Journal
Journal of Cardiac Arrhythmias
Published
2025-03-24
DOI
https://doi.org/10.24207/jca.v38i1.3561
Primary Topic
Cardiac pacing and defibrillation studies
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Early displacement of an atrial lead: A fatal presentation of a relatively frequent complication

Antoni Martínez‐Rubio, Jordi Puntí, Soledad Ascoeta, Gabriel Gusi et al.
Journal of Cardiac Arrhythmias
Cardiac pacing and defibrillation studies
article

Early displacement of an atrial lead: A fatal presentation of a relatively frequent complication

Antoni Martínez‐Rubio, Jordi Puntí, Soledad Ascoeta, Gabriel Gusi, Victor García-Hernando, Gala Caixal
article en

Abstract

An 81-year-old woman was referred to Parc Taulí Hospital Universitari with complete atrioventricular block and a dual-chamber pacemaker was implanted. She subsequently presented an episode of ventricular tachycardia that degenerated into ventricular fibrillation requiring cardiopulmonary resuscitation. After review of the device, as well as the electrocardiogram and chest X-ray, displacement of the atrial lead into the right ventricle was confirmed. Contrary to expectations, the ventricular event was not secondary to endocardial irritation by the displaced lead, so the hypothesis of a possible depolarization-repolarization heterogeneity as a cause of the increased arrhythmogenicity was raised.

Journal of Cardiac ArrhythmiasVol. 38
Universitat Autònoma de Barcelona (ES), Hospital de Sabadell (ES)
Openalex Percentile: Top 92%
Cardiac pacing and defibrillation studies
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.