Evaluation of the system accuracy of frameless stereotactic radiosurgery using a combination of cone beam CT, six degrees of freedom couch, and surface image‐guided systems
This study introduces an innovative approach to quantifying the impact of couch rotation on the FSRS system using head phantoms with turntables. The overall accuracy of the FSRS system was on the order of 1.1 mm/1°.
Authors
- Chihray Liu
- Sheng‐Hsuan Sun
- Jonathan Li (ORCID: https://orcid.org/0000-0002-0916-6678)
- Guanghua Yan
- Hao‐Wen Cheng
Institutions
- University of Florida (US)
Publication Details
- Journal
- Journal of Applied Clinical Medical Physics
- Published
- 2025-03-22
- DOI
- https://doi.org/10.1002/acm2.70082
- Primary Topic
- Advanced Radiotherapy Techniques
- Type
- article
- Field-Weighted Citation Impact
- 0.00