Evaluation of the system accuracy of frameless stereotactic radiosurgery using a combination of cone beam CT, six degrees of freedom couch, and surface image‐guided systems

This study introduces an innovative approach to quantifying the impact of couch rotation on the FSRS system using head phantoms with turntables. The overall accuracy of the FSRS system was on the order of 1.1 mm/1°.

Authors

Institutions

Publication Details

Journal
Journal of Applied Clinical Medical Physics
Published
2025-03-22
DOI
https://doi.org/10.1002/acm2.70082
Primary Topic
Advanced Radiotherapy Techniques
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Evaluation of the system accuracy of frameless stereotactic radiosurgery using a combination of cone beam CT, six degrees of freedom couch, and surface image‐guided systems

Chihray Liu, Sheng‐Hsuan Sun, Jonathan Li, Guanghua Yan et al.
2 links
Journal of Applied Clinical Medical Physics
Advanced Radiotherapy Techniques
article

Evaluation of the system accuracy of frameless stereotactic radiosurgery using a combination of cone beam CT, six degrees of freedom couch, and surface image‐guided systems

Chihray Liu, Sheng‐Hsuan Sun, Jonathan Li, Guanghua Yan, Hao‐Wen Cheng
article en

Abstract

This study introduces an innovative approach to quantifying the impact of couch rotation on the FSRS system using head phantoms with turntables. The overall accuracy of the FSRS system was on the order of 1.1 mm/1°.

Journal of Applied Clinical Medical PhysicsVol. 26(6)
University of Florida (US)
Peace, Justice and strong institutions
Openalex Percentile: Top 93%
Advanced Radiotherapy Techniques
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.