Validation of a High‐Performance Liquid Chromatography Separation of Weak Acids, Neutral Compounds, and Permanent Ions Using Sequential Elution with Tandem‐Column Liquid Chromatography
Sequential elution-liquid chromatography (SE-LC) employs conventional high-performance LC instrumentation to separate compounds by class and within each class using two or more sequential elution modes. The aim of this work was to validate a novel SE-LC method used to separate weak acids, permanently charged ions (anions), and neutral compounds by applying first a low pH isocratic elution to elute the weak acids, then an acetonitrile gradient to elute the neutral compounds, and lastly a sodium methanesulfonate gradient to elute the anionic compounds using a tandem-column arrangement comprised of a C18 column and a strong anion exchange column. The effects of flow rate, starting mobile phase organic content, column temperature, mobile phase composition, and gradient start time were investigated using a 2<sup>5-1</sup> fractional factorial design with center points for robustness. The method was demonstrated to be robust, specific, linear, accurate, and precise (repeatable) over the range of 6%-120% of the target analyte concentrations.
Authors
- Kaitlin M. Grinias
- Joe P. Foley (ORCID: https://orcid.org/0000-0002-8388-6112)
- Lauren Kline Lovejoy
Institutions
- GlaxoSmithKline (United States) (US)
- Kalamazoo College (US)
- Drexel University (US)
Publication Details
- Journal
- Journal of Separation Science
- Published
- 2025-03-01
- DOI
- https://doi.org/10.1002/jssc.70124
- Primary Topic
- Analytical Chemistry and Chromatography
- Type
- article
- Field-Weighted Citation Impact
- 0.00