Quantitative estimation of blood loss volume using laser speckle flow index signals from a wireless wearable device (Conference Presentation)

Postpartum hemorrhage (PPH) is the leading and most preventable cause of maternal mortality. There is an urgent need for a technology that can monitor visible and occult blood loss in real-time to detect PPH in early stages. A compact, low-cost (<$150), wrist worn wearable device that continuously monitors perfusion via laser speckle flow index (LSFI) was previously developed to detect blood loss. The device was used to captures pulsatile LSFI waveforms from swine subjects during blood removal. Collected waveforms were used to extract a set of features that describe the waveform shape, and the features were used to train a multivariate regression model to predict blood loss volume. The average absolute error was 137 mL, where the correlation between true and predicted blood loss volume was r=0.86. When hemoglobin concentration was also included as a feature in the model, the average error was reduced to 60.35 mL and r=0.93.

Authors

Institutions

Publication Details

Published
2025-03-20
DOI
https://doi.org/10.1117/12.3040929
Primary Topic
Non-Invasive Vital Sign Monitoring
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Quantitative estimation of blood loss volume using laser speckle flow index signals from a wireless wearable device (Conference Presentation)

C. O’Brien, Madison Carlgren, Peinan Zhao, Toi Collins et al.
Non-Invasive Vital Sign Monitoring
article

Quantitative estimation of blood loss volume using laser speckle flow index signals from a wireless wearable device (Conference Presentation)

C. O’Brien, Madison Carlgren, Peinan Zhao, Toi Collins, Leonid Shmuylovich, Todd Pavek, Francesca Bonetta-Misteli, Antonina I. Frolova
article en

Abstract

Postpartum hemorrhage (PPH) is the leading and most preventable cause of maternal mortality. There is an urgent need for a technology that can monitor visible and occult blood loss in real-time to detect PPH in early stages. A compact, low-cost (<$150), wrist worn wearable device that continuously monitors perfusion via laser speckle flow index (LSFI) was previously developed to detect blood loss. The device was used to captures pulsatile LSFI waveforms from swine subjects during blood removal. Collected waveforms were used to extract a set of features that describe the waveform shape, and the features were used to train a multivariate regression model to predict blood loss volume. The average absolute error was 137 mL, where the correlation between true and predicted blood loss volume was r=0.86. When hemoglobin concentration was also included as a feature in the model, the average error was reduced to 60.35 mL and r=0.93.

Washington University in St. Louis (US)
Openalex Percentile: Top 95%
Non-Invasive Vital Sign Monitoring
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.