Long-term stability from adaptive autocalibration in mid-infrared sensing instruments
Mid-infrared laser-based sensing is widely used for the detection and quantification of trace gases for environmental monitoring, industrial process control, defense, and security applications. However, such laser spectroscopy-based field instruments require frequent calibration to correct for long-term drifts and precision. In many situations, such calibrations may be redundant and may not be effective in monitoring instrument errors or discriminating “real” environmental effects with instrument drifts. This paper shows that physics and diffusion-based models can be developed to predict gradual or rapid changes in density at given temperatures and appropriate source characterizations. We used a 3.3μm, laser-based instrument to demonstrate an adaptive and auto-calibration methodology to account for drifts during field deployments and an adaptive method to validate environmental changes, e.g., plume detection, diurnal variations, and flux emissions. We field operation and implementation of adaptive auto-calibration. The instruments are used for the detection of methane and water vapor, but an adaptive auto-calibration technique can be employed for any laser sensing instrument.
Authors
- Al R. Alexis
- Lauren Anderson (ORCID: https://orcid.org/0000-0001-5725-9329)
- M. Amir Khan (ORCID: https://orcid.org/0000-0003-0143-4161)
Institutions
- Delaware State University (US)
Publication Details
- Published
- 2025-03-20
- DOI
- https://doi.org/10.1117/12.3042496
- Primary Topic
- Calibration and Measurement Techniques
- Type
- article
- Field-Weighted Citation Impact
- 0.00