Laser coagulation of blood vessels at 454 nm wavelength for neurosurgical interventions: occlusion rates, process dynamics, and thermal damage

In neurosurgery, where operations take place near tissue structures with high functionality, precise devices for microsurgical procedures such as blood vessel coagulation are crucial. Currently, bipolar forceps that deliver up to 60W with high alternating current are used for vascular coagulation (hemostasis) to thermally seal blood vessels and stop bleeding. However, the high current can disturb electrophysiological monitoring and cause nerve damage from heat spread. Our approach uses a 454 nm laser diode, which closely matches the hemoglobin absorption peak, to directly heat the blood and avoid thermal damage to surrounding tissue. In experiments on blood vessels at the vascular tree of pig hearts, occlusion rates at different irradiation durations, the temperature of the adjacent tissue, and the dynamics of the coagulation process using optical coherence tomography (OCT) were investigated. Our findings show that a laser radiation of 454 nm wavelength can reliably coagulate vessels up to 400 μm in diameter, offering a safe microsurgical method for neurosurgery.

Authors

Publication Details

Published
2025-03-20
DOI
https://doi.org/10.1117/12.3040167
Primary Topic
Laser-Ablation Synthesis of Nanoparticles
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Laser coagulation of blood vessels at 454 nm wavelength for neurosurgical interventions: occlusion rates, process dynamics, and thermal damage

Achim Lenenbach, Christina Giesen, Lazar Bochvarov, Elisa Jarry
Laser-Ablation Synthesis of Nanoparticles
article

Laser coagulation of blood vessels at 454 nm wavelength for neurosurgical interventions: occlusion rates, process dynamics, and thermal damage

Achim Lenenbach, Christina Giesen, Lazar Bochvarov, Elisa Jarry
article en

Abstract

In neurosurgery, where operations take place near tissue structures with high functionality, precise devices for microsurgical procedures such as blood vessel coagulation are crucial. Currently, bipolar forceps that deliver up to 60W with high alternating current are used for vascular coagulation (hemostasis) to thermally seal blood vessels and stop bleeding. However, the high current can disturb electrophysiological monitoring and cause nerve damage from heat spread. Our approach uses a 454 nm laser diode, which closely matches the hemoglobin absorption peak, to directly heat the blood and avoid thermal damage to surrounding tissue. In experiments on blood vessels at the vascular tree of pig hearts, occlusion rates at different irradiation durations, the temperature of the adjacent tissue, and the dynamics of the coagulation process using optical coherence tomography (OCT) were investigated. Our findings show that a laser radiation of 454 nm wavelength can reliably coagulate vessels up to 400 μm in diameter, offering a safe microsurgical method for neurosurgery.

Openalex Percentile: Top 95%
Laser-Ablation Synthesis of Nanoparticles
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Laser coagulation of blood vessels at 454 nm wavelength for neurosurgical interventions: occlusion rates, process dynamics, and thermal damage — Achim Lenenbach, Christina Giesen, et al. · (2025) | TGRS Research Map | TGRS