Accuracy and repeatability of the COSMED® Q-NRG max mobile metabolic system
The Q-NRG Max® is a valid and reliable mobile metabolic system for the measurement of ventilation, oxygen consumption, and carbon dioxide production. Measurements were below the 5% TE and MDC, and 2% MAPE recommended thresholds across a wide range of metabolic rates up to 6 L/min oxygen consumption.
Authors
- Carlo Baldari (ORCID: https://orcid.org/0000-0001-6227-3571)
- Marco Meucci (ORCID: https://orcid.org/0000-0002-1549-2082)
- Lavinia Falcioni (ORCID: https://orcid.org/0000-0001-8704-7044)
- Laura Guidetti (ORCID: https://orcid.org/0000-0002-1419-4550)
- Andrey Sanko Posada
- Chris Wing
- Luke Dover
Institutions
- Università degli Studi eCampus (IT)
- University Niccolò Cusano (IT)
- Appalachian State University (US)
Publication Details
- Journal
- PLoS ONE
- Published
- 2025-03-20
- DOI
- https://doi.org/10.1371/journal.pone.0319394
- Citations
- 2
- Primary Topic
- Cardiovascular and exercise physiology
- Type
- article
- Field-Weighted Citation Impact
- 4.76