Assessment of [<sup>177</sup>Lu]Lu-DOTATATE Dosimetry from High-Speed Whole-Body Recordings Provided by a 360° Cadmium–Zinc–Telluride Camera Compared with Results from a Conventional Anger-Camera Protocol

Absorbed doses (ADs) may be calculated through serial conventional SPECT imaging after therapeutic [<sup>177</sup>Lu]Lu-DOTATATE injection but with recording times too long for clinical routine. The aim of this study was to determine whether activity concentrations and ADs calculated from a high-speed whole-body 360° cadmium-zinc-telluride (CZT) SPECT camera are comparable to those provided by conventional SPECT. <b>Methods:</b> Fifteen patients referred for [<sup>177</sup>Lu]Lu-DOTATATE treatment were enrolled and underwent, at 24, 96, and 168 h after [<sup>177</sup>Lu]Lu-DOTATATE injection, 32 min of thoracoabdominopelvic conventional SPECT recording and 18 min of whole-body CZT SPECT recording. The order of device procedure (conventional or CZT SPECT) was randomly selected to avoid bias. The kidneys, bone marrow, liver, and spleen were automatically segmented by an artificial intelligence-based tool on CT images, whereas the largest tumor was segmented manually on CT or MR images. Time-integrated activity curves were computed using activity concentrations from the kidneys, bone marrow, liver, spleen, and tumor and convolved by the voxel S-value kernel to calculate AD values. The partial-volume effect was corrected by fitting the recovery coefficients of 1- to 113-mL spheres. Activity concentrations (in Bq/mL) and ADs (in Gy/GBq) were expressed as median (minimum to maximum) and compared between the 2 cameras with Wilcoxon tests. <b>Results:</b> Activity concentrations and ADs were not significantly different between conventional and CZT SPECT, respectively, for the kidneys (337.3 [203.3-622.4] vs. 342.7 [218.2-547.8] kBq/mL at 24 h and 0.45 [0.26-0.79] vs. 0.47 [0.25-0.79] Gy/GBq), spleen (340.3 [171.8-703.1] vs. 277.7 [215.6-640.8] kBq/mL at 24 h and 0.48 [0.31-1.41] vs. 0.50 [0.35-1.15] Gy/GBq), bone marrow (32.7 [13.9-662.9] vs. 41.1 [12.4-509.6] kBq/mL at 24 h and 0.09 [0.04-1.96] vs. 0.13 [0.06-0.87] Gy/GBq), and largest tumor (826.8 [376.4-4,459.0] vs. 1,116.7 [447.0-6,043.3] kBq/mL at 24 h and 1.8 [0.9-5.7] vs. 2.0 [1.0-7.3] Gy/GBq). However, significant, albeit small, differences were observed for the liver (126.2 [42.7-335.9] vs. 113.5 [35.1-293.4] kBq/mL at 24 h and 0.23 [0.07-0.72] vs. 0.21 [0.05-0.64] Gy/GBq, both <i>P</i> < 0.01). <b>Conclusion:</b> Activity concentrations and absorbed doses calculated from the 360° CZT SPECT/CT system after [<sup>177</sup>Lu]Lu-DOTATATE injection are globally comparable to those obtained from a conventional SPECT/CT system. However, CZT SPECT/CT is more appropriate for clinical routine, with shorter recording times and actual whole-body coverage.

Authors

Institutions

Publication Details

Journal
Journal of Nuclear Medicine
Published
2025-03-20
DOI
https://doi.org/10.2967/jnumed.124.268910
Citations
6
Primary Topic
Medical Imaging Techniques and Applications
Type
article
Field-Weighted Citation Impact
8.16
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Assessment of [177Lu]Lu-DOTATATE Dosimetry from High-Speed Whole-Body Recordings Provided by a 360° Cadmium–Zinc–Telluride Camera Compared with Results from a Conventional Anger-Camera Protocol

Quentin Citerne, Gabriela Hossu, Timothée Zaragori, Laëtitia Imbert et al.
6 citations
Journal of Nuclear Medicine
Medical Imaging Techniques and Applications
8.16
article

Assessment of [177Lu]Lu-DOTATATE Dosimetry from High-Speed Whole-Body Recordings Provided by a 360° Cadmium–Zinc–Telluride Camera Compared with Results from a Conventional Anger-Camera Protocol

Quentin Citerne, Gabriela Hossu, Timothée Zaragori, Laëtitia Imbert, Caroline Boursier, Véronique Roch, Pierre-Yves Marie, Elodie Chevalier
article en
6 citations

Abstract

Absorbed doses (ADs) may be calculated through serial conventional SPECT imaging after therapeutic [<sup>177</sup>Lu]Lu-DOTATATE injection but with recording times too long for clinical routine. The aim of this study was to determine whether activity concentrations and ADs calculated from a high-speed whole-body 360° cadmium-zinc-telluride (CZT) SPECT camera are comparable to those provided by conventional SPECT. <b>Methods:</b> Fifteen patients referred for [<sup>177</sup>Lu]Lu-DOTATATE treatment were enrolled and underwent, at 24, 96, and 168 h after [<sup>177</sup>Lu]Lu-DOTATATE injection, 32 min of thoracoabdominopelvic conventional SPECT recording and 18 min of whole-body CZT SPECT recording. The order of device procedure (conventional or CZT SPECT) was randomly selected to avoid bias. The kidneys, bone marrow, liver, and spleen were automatically segmented by an artificial intelligence-based tool on CT images, whereas the largest tumor was segmented manually on CT or MR images. Time-integrated activity curves were computed using activity concentrations from the kidneys, bone marrow, liver, spleen, and tumor and convolved by the voxel S-value kernel to calculate AD values. The partial-volume effect was corrected by fitting the recovery coefficients of 1- to 113-mL spheres. Activity concentrations (in Bq/mL) and ADs (in Gy/GBq) were expressed as median (minimum to maximum) and compared between the 2 cameras with Wilcoxon tests. <b>Results:</b> Activity concentrations and ADs were not significantly different between conventional and CZT SPECT, respectively, for the kidneys (337.3 [203.3-622.4] vs. 342.7 [218.2-547.8] kBq/mL at 24 h and 0.45 [0.26-0.79] vs. 0.47 [0.25-0.79] Gy/GBq), spleen (340.3 [171.8-703.1] vs. 277.7 [215.6-640.8] kBq/mL at 24 h and 0.48 [0.31-1.41] vs. 0.50 [0.35-1.15] Gy/GBq), bone marrow (32.7 [13.9-662.9] vs. 41.1 [12.4-509.6] kBq/mL at 24 h and 0.09 [0.04-1.96] vs. 0.13 [0.06-0.87] Gy/GBq), and largest tumor (826.8 [376.4-4,459.0] vs. 1,116.7 [447.0-6,043.3] kBq/mL at 24 h and 1.8 [0.9-5.7] vs. 2.0 [1.0-7.3] Gy/GBq). However, significant, albeit small, differences were observed for the liver (126.2 [42.7-335.9] vs. 113.5 [35.1-293.4] kBq/mL at 24 h and 0.23 [0.07-0.72] vs. 0.21 [0.05-0.64] Gy/GBq, both <i>P</i> < 0.01). <b>Conclusion:</b> Activity concentrations and absorbed doses calculated from the 360° CZT SPECT/CT system after [<sup>177</sup>Lu]Lu-DOTATATE injection are globally comparable to those obtained from a conventional SPECT/CT system. However, CZT SPECT/CT is more appropriate for clinical routine, with shorter recording times and actual whole-body coverage.

Journal of Nuclear MedicineVol. 66(5)
Inserm (FR), Centre d’Investigation Clinique Innovation Technologique de Nancy (FR), Imagerie Adaptative Diagnostique et Interventionnelle (FR), Université de Lorraine (FR)
Openalex Percentile: Top 3%
Medical Imaging Techniques and Applications
8.16
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.