Effectiveness of Radiation Shields to Minimize Operator Dose in the Bronchoscopy Suite: A Phantom Study and Clinical Application

<b>Background/Objectives</b>: Fluoroscopy has been widely adopted in interventional pulmonology, as it facilitates real-time visualization of the bronchoscope, endobronchial ultrasound, and biopsy tools during procedures. The purpose of this study was to evaluate the effectiveness of radiation shields in minimizing scattered X-ray dose to the bronchoscopist in a phantom study and to determine the dose of scattered X-ray dose to medical staff with radiation shields in clinical application. <b>Methods</b>: An anthropomorphic torso phantom was positioned on the fluoroscopic table between the C-arm X-ray tube and the image detector to mimic bronchoscopic operations. Upper and lower body lead shields were used to examine the effectiveness of radiation shielding. Scatter radiation rates were assessed at a first operator location using real-time dosimeters with and without protective devices. In clinical application, the scattered X-ray dose of the first operator and main assistant was measured using wearable radiation dosimeters during 20 procedures. <b>Results</b>: In the phantom study, scattered radiation without shielding was 266.34 ± 8.86 μSv/h (glabella), 483.90 ± 8.01 μSv/h (upper thorax), 143.97 ± 8.20 μSv/h (hypogastrium), and 7.22 ± 0.28 μSv/h (ankle). The combination of upper and lower body lead shields reduced the scattered X-ray dose by 98.7%, 98.3%, 66.2%, and 79.9% at these levels, respectively. In clinical application, mean scattered X-ray dose rates were 0.14 ± 0.05 μSv/procedure (eye), 0.46 ± 0.51 μSv/procedure (chest), 0.67 ± 0.50 μSv/procedure (hypogastrium), and 1.57 ± 2.84 μSv/procedure (assistant's wrist). <b>Conclusions</b>: The combination of radiation shields significantly reduced the scattered X-ray dose at the operator site in the phantom study. The scattered X-ray dose to medical staff during bronchoscopy can be kept at a low level with the aid of a shielding system.

Authors

Institutions

Publication Details

Journal
Journal of Clinical Medicine
Published
2025-03-20
DOI
https://doi.org/10.3390/jcm14062114
Citations
1
Primary Topic
Radiation Dose and Imaging
Type
article
Field-Weighted Citation Impact
1.36
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Effectiveness of Radiation Shields to Minimize Operator Dose in the Bronchoscopy Suite: A Phantom Study and Clinical Application

Won Chul Shin, Suk‐Woong Kang, Hee Yun Seol, Taehoon Lee et al.
1 citations
Journal of Clinical Medicine
Radiation Dose and Imaging
1.36
article

Effectiveness of Radiation Shields to Minimize Operator Dose in the Bronchoscopy Suite: A Phantom Study and Clinical Application

Won Chul Shin, Suk‐Woong Kang, Hee Yun Seol, Taehoon Lee, Hyun Sung Chung, Ji Hyeon Joo, Yongkan Ki, Hosang Jeon, Seong Hoon Yoon, Yun Seong Kim, Seung Hyun Yong, Dong Woon Kim
article en
1 citations

Abstract

<b>Background/Objectives</b>: Fluoroscopy has been widely adopted in interventional pulmonology, as it facilitates real-time visualization of the bronchoscope, endobronchial ultrasound, and biopsy tools during procedures. The purpose of this study was to evaluate the effectiveness of radiation shields in minimizing scattered X-ray dose to the bronchoscopist in a phantom study and to determine the dose of scattered X-ray dose to medical staff with radiation shields in clinical application. <b>Methods</b>: An anthropomorphic torso phantom was positioned on the fluoroscopic table between the C-arm X-ray tube and the image detector to mimic bronchoscopic operations. Upper and lower body lead shields were used to examine the effectiveness of radiation shielding. Scatter radiation rates were assessed at a first operator location using real-time dosimeters with and without protective devices. In clinical application, the scattered X-ray dose of the first operator and main assistant was measured using wearable radiation dosimeters during 20 procedures. <b>Results</b>: In the phantom study, scattered radiation without shielding was 266.34 ± 8.86 μSv/h (glabella), 483.90 ± 8.01 μSv/h (upper thorax), 143.97 ± 8.20 μSv/h (hypogastrium), and 7.22 ± 0.28 μSv/h (ankle). The combination of upper and lower body lead shields reduced the scattered X-ray dose by 98.7%, 98.3%, 66.2%, and 79.9% at these levels, respectively. In clinical application, mean scattered X-ray dose rates were 0.14 ± 0.05 μSv/procedure (eye), 0.46 ± 0.51 μSv/procedure (chest), 0.67 ± 0.50 μSv/procedure (hypogastrium), and 1.57 ± 2.84 μSv/procedure (assistant's wrist). <b>Conclusions</b>: The combination of radiation shields significantly reduced the scattered X-ray dose at the operator site in the phantom study. The scattered X-ray dose to medical staff during bronchoscopy can be kept at a low level with the aid of a shielding system.

Journal of Clinical MedicineVol. 14(6)
Ulsan College (KR), Yonsei University (KR), Severance Hospital (KR), University of Ulsan (KR), National Cancer Center (KR), Ulsan University Hospital (KR), Pusan National University Yangsan Hospital (KR), The University of Texas Southwestern Medical Center (US)
Good health and well-being
Openalex Percentile: Top 22%
Radiation Dose and Imaging
1.36
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.