The AVEIR DR (Abbott Medical) was the first dual-chamber leadless pacing system approved by the US Food and Drug Administration. Many patients with sinus node dysfunction have physiologically normal atrioventricular conduction, so an atrial-only pacing system may be appropriate for them. Three patients with symptomatic sinus node dysfunction underwent device implantation with electrophysiologic study, including incremental atrial pacing until loss of 1:1 atrioventricular conduction (atrioventricular Wenckebach cycle length) and His-ventricular interval measurement. If the atrioventricular Wenckebach cycle length was no more than 400 milliseconds and the His-ventricular interval measurement was no more than 55 milliseconds, single-chamber atrial implantation was deemed appropriate. Each patient displayed a different response to atrioventricular conduction testing, demonstrating how electrophysiologic study before device implantation may identify patients for whom atrial-only pacing is not appropriate.

Authors

Institutions

Publication Details

Journal
Texas Heart Institute Journal
Published
2025-01-01
DOI
https://doi.org/10.14503/thij-24-8495
Citations
1
Primary Topic
Cardiac pacing and defibrillation studies
Type
article
Field-Weighted Citation Impact
1.39
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

The Role of Electrophysiologic Study in Device Selection for Leadless Atrial Pacing

Joshua R. Hirsch, Jordan M. Gales, Hamid Afshar, Irakli Giorgberidzė
1 citations
Texas Heart Institute Journal
Cardiac pacing and defibrillation studies
1.39
article

The Role of Electrophysiologic Study in Device Selection for Leadless Atrial Pacing

Joshua R. Hirsch, Jordan M. Gales, Hamid Afshar, Irakli Giorgberidzė
article en
1 citations

Abstract

The AVEIR DR (Abbott Medical) was the first dual-chamber leadless pacing system approved by the US Food and Drug Administration. Many patients with sinus node dysfunction have physiologically normal atrioventricular conduction, so an atrial-only pacing system may be appropriate for them. Three patients with symptomatic sinus node dysfunction underwent device implantation with electrophysiologic study, including incremental atrial pacing until loss of 1:1 atrioventricular conduction (atrioventricular Wenckebach cycle length) and His-ventricular interval measurement. If the atrioventricular Wenckebach cycle length was no more than 400 milliseconds and the His-ventricular interval measurement was no more than 55 milliseconds, single-chamber atrial implantation was deemed appropriate. Each patient displayed a different response to atrioventricular conduction testing, demonstrating how electrophysiologic study before device implantation may identify patients for whom atrial-only pacing is not appropriate.

Texas Heart Institute JournalVol. 52(1)
Michael E. DeBakey VA Medical Center (US), Baylor College of Medicine (US), St. Luke's Hospital (US)
Zero hunger
Openalex Percentile: Top 21%
Cardiac pacing and defibrillation studies
1.39
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.