Determination of confocal parameters of OCT imaging for eliminating confocal effect on attenuation coefficient estimation

Optical coherence tomography (OCT) provides three-dimensional images of biological tissues based on low-coherence optical interference. Attenuation coefficient estimation is one of the important functional extensions of OCT and has received many research efforts as attenuation coefficient has been found to be associated with histopathological transformation of human tissues. However, attenuation coefficient estimation accuracy is deteriorated by the confocal effect on OCT A-scan. Thus, it is desired to eliminate the confocal effect, which requires accurate determination of the confocal parameters. In this study, we propose what we believe to be a novel method for confocal parameter extraction, called dual NA ratio fitting (DNRF). DNRF requires a repetition of B-scan with varied numerical apertures (NA), altering the Rayleigh length of B-scan but keeping the focal depth fixed. Then, the focal depth and Rayleigh length can be determined by fitting an A-scan ratio function from the two repeated B-scans. The NA tuning was achieved by adding a beam expansion module into the sample arm. The method was evaluated on intralipid samples and multi-layer phantoms. The results demonstrate that our method is capable of determine the confocal parameters with good accuracy. With the extracted confocal parameters, the confocal effect was removed effectively, upon which attenuation coefficient estimation using traditional depth-resolved method appeared to be more accurate than the confocal parameter extraction based on the A-scan ratio function of two repeated B-scans with varied focal depths. Last, our method was tested on human skin <i>in vivo</i>, yielding attenuation coefficients consistent with literature. This indicates good potential of our method to be used for clinical applications.

Authors

Institutions

Publication Details

Journal
Optics Express
Published
2025-03-19
DOI
https://doi.org/10.1364/oe.555753
Primary Topic
Optical Coherence Tomography Applications
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Determination of confocal parameters of OCT imaging for eliminating confocal effect on attenuation coefficient estimation

Guohua Shi, Xinyuan Cao, Jianhua Mo, Zhilong Yan
Optics Express
Optical Coherence Tomography Applications
article

Determination of confocal parameters of OCT imaging for eliminating confocal effect on attenuation coefficient estimation

Guohua Shi, Xinyuan Cao, Jianhua Mo, Zhilong Yan
article en

Abstract

Optical coherence tomography (OCT) provides three-dimensional images of biological tissues based on low-coherence optical interference. Attenuation coefficient estimation is one of the important functional extensions of OCT and has received many research efforts as attenuation coefficient has been found to be associated with histopathological transformation of human tissues. However, attenuation coefficient estimation accuracy is deteriorated by the confocal effect on OCT A-scan. Thus, it is desired to eliminate the confocal effect, which requires accurate determination of the confocal parameters. In this study, we propose what we believe to be a novel method for confocal parameter extraction, called dual NA ratio fitting (DNRF). DNRF requires a repetition of B-scan with varied numerical apertures (NA), altering the Rayleigh length of B-scan but keeping the focal depth fixed. Then, the focal depth and Rayleigh length can be determined by fitting an A-scan ratio function from the two repeated B-scans. The NA tuning was achieved by adding a beam expansion module into the sample arm. The method was evaluated on intralipid samples and multi-layer phantoms. The results demonstrate that our method is capable of determine the confocal parameters with good accuracy. With the extracted confocal parameters, the confocal effect was removed effectively, upon which attenuation coefficient estimation using traditional depth-resolved method appeared to be more accurate than the confocal parameter extraction based on the A-scan ratio function of two repeated B-scans with varied focal depths. Last, our method was tested on human skin <i>in vivo</i>, yielding attenuation coefficients consistent with literature. This indicates good potential of our method to be used for clinical applications.

Optics ExpressVol. 33(7)
Suzhou Institute of Biomedical Engineering and Technology (CN)
National Key Research and Development Program of China
Openalex Percentile: Top 95%
Optical Coherence Tomography Applications
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.