Assessment of a Patient Dose Monitoring System for Average Glandular Dose (AGD) Estimate in Mammography

This study assessed the accuracy of average glandular dose (AGD) calculations for two Selenia Dimensions mammography systems using data from the online dose management DoseWatch software version 3.3.5.1. Mammographic images acquired between January 2021 and December 2022 were retrospectively analyzed. The AGD values displayed by the systems were compared with those independently calculated using the Dance and Boone methods. Additionally, real glandular composition of breast was estimated using LIBRA (Laboratory for Individualized Breast Radiodensity Assessment) software version 1.0.4 for a selected subgroup of patients. Results showed that the AGD values displayed by the systems were generally consistent with those calculated using the Dance method, but discrepancies emerged when applying the Boone method, especially when using estimated glandular composition. Most mammograms fell within acceptable and achievable dose limits according to European guidelines, though a small percentage exceeded these thresholds. The findings suggest that the Dance method, using glandular composition estimated through LIBRA, provides a reliable and accurate AGD calculation, offering a simpler alternative to more complex individualized calculations. The study highlights the importance of accurate glandularity estimation for proper dose management in mammography.

Authors

Institutions

Publication Details

Journal
Applied Sciences
Published
2025-03-19
DOI
https://doi.org/10.3390/app15063338
Citations
2
Primary Topic
Digital Radiography and Breast Imaging
Type
article
Field-Weighted Citation Impact
3.04
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Assessment of a Patient Dose Monitoring System for Average Glandular Dose (AGD) Estimate in Mammography

C. Marino, Alessia D’Anna, Giuseppe Stella, Martina Pace et al.
2 citations
Applied Sciences
Digital Radiography and Breast Imaging
3.04
article

Assessment of a Patient Dose Monitoring System for Average Glandular Dose (AGD) Estimate in Mammography

C. Marino, Alessia D’Anna, Giuseppe Stella, Martina Pace, G. Borzì, Lucia Zirone, E. Bonanno, N. Cavalli
article en
2 citations

Abstract

This study assessed the accuracy of average glandular dose (AGD) calculations for two Selenia Dimensions mammography systems using data from the online dose management DoseWatch software version 3.3.5.1. Mammographic images acquired between January 2021 and December 2022 were retrospectively analyzed. The AGD values displayed by the systems were compared with those independently calculated using the Dance and Boone methods. Additionally, real glandular composition of breast was estimated using LIBRA (Laboratory for Individualized Breast Radiodensity Assessment) software version 1.0.4 for a selected subgroup of patients. Results showed that the AGD values displayed by the systems were generally consistent with those calculated using the Dance method, but discrepancies emerged when applying the Boone method, especially when using estimated glandular composition. Most mammograms fell within acceptable and achievable dose limits according to European guidelines, though a small percentage exceeded these thresholds. The findings suggest that the Dance method, using glandular composition estimated through LIBRA, provides a reliable and accurate AGD calculation, offering a simpler alternative to more complex individualized calculations. The study highlights the importance of accurate glandularity estimation for proper dose management in mammography.

Applied SciencesVol. 15(6)
University of Catania (IT), Fondazione Humanitas per la Ricerca (IT)
Openalex Percentile: Top 10%
Digital Radiography and Breast Imaging
3.04
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Assessment of a Patient Dose Monitoring System for Average Glandular Dose (AGD) Estimate in Mammography — C. Marino, Alessia D’Anna, et al. · Applied Sciences (2025) | TGRS Research Map | TGRS