Comparison of mechanical resuscitation by an LV Impella device to extracorporeal resuscitation using VAECMO in a large animal model

Extracorporeal cardiopulmonary resuscitation (ECPR) is an effective treatment for cardiac arrest (CA). Percutaneous left ventricular (LV) assist devices such as the Impella ECP (intravascular CPR [ICPR]) have been proposed as a less invasive alternative. The aim of this study was to explore the haemodynamic differences between ECPR and ICPR using a large animal model of electrically induced CA. Fourteen juvenile female German landrace pigs (72.4 ± 9.8 kg) were subjected to electrically induced CA for 5 mins followed by either ECPR (veno-arterial extracorporeal membrane oxygenation [VA-ECMO]) or ICPR (Impella ECP). Haemodynamic parameters and echocardiographic ventricular function indicators were monitored. Mechanical circulatory support (MCS) was continued until five hours after the return of spontaneous circulation (ROSC), when the devices were removed. Resuscitation outcomes and the haemodynamic effects of ECPR and ICPR were compared. The cannulation time for ECMO (469 ± 129 s) was significantly longer than the time for Impella device implantation (153 ± 64 s, p < 0.001). ECPR facilitated ROSC in 6/6 animals, whereas ICPR facilitated ROSC in 6/8 animals (p = 0.19). Echocardiography revealed no difference in LV or right ventricular (RV) dysfunction between the ECPR- and ICPR-treated animals after resuscitation (LV-global longitudinal strain [GLS] 3 h post-ROSC: ICPR: - 16.5 ± 5.6% vs. ECPR: - 13.7 ± 5.9%, p = 0.99; RV-GLS 3 h post-ROSC: ICPR: - 15.9 ± 3.3% vs. ECPR: - 17.3 ± 10.6%, p = 0.99). MCS using VA-ECMO and the Impella device both provided effective haemodynamic support during CA and post-ROSC in this large animal model. Despite LV unloading conferring a hypothetical advantage for ICPR, no significant differences in myocardial recovery were observed.

Authors

Institutions

Publication Details

Journal
Scientific Reports
Published
2025-03-19
DOI
https://doi.org/10.1038/s41598-025-93264-2
Citations
1
Primary Topic
Cardiac Arrest and Resuscitation
Type
article
Field-Weighted Citation Impact
2.25

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Comparison of mechanical resuscitation by an LV Impella device to extracorporeal resuscitation using VAECMO in a large animal model

Sebastian Billig, Nadine Hochhausen, Matthias Derwall, Katharina Nubbemeyer et al.
1 citations
Scientific Reports
Cardiac Arrest and Resuscitation
2.25
article

Comparison of mechanical resuscitation by an LV Impella device to extracorporeal resuscitation using VAECMO in a large animal model

Sebastian Billig, Nadine Hochhausen, Matthias Derwall, Katharina Nubbemeyer, Alexander Theißen, Christoph Nix, Eveline Bennek-Schoepping, Siarhei Yelenski, Adomas Kanauskas
article en
1 citations

Abstract

Extracorporeal cardiopulmonary resuscitation (ECPR) is an effective treatment for cardiac arrest (CA). Percutaneous left ventricular (LV) assist devices such as the Impella ECP (intravascular CPR [ICPR]) have been proposed as a less invasive alternative. The aim of this study was to explore the haemodynamic differences between ECPR and ICPR using a large animal model of electrically induced CA. Fourteen juvenile female German landrace pigs (72.4 ± 9.8 kg) were subjected to electrically induced CA for 5 mins followed by either ECPR (veno-arterial extracorporeal membrane oxygenation [VA-ECMO]) or ICPR (Impella ECP). Haemodynamic parameters and echocardiographic ventricular function indicators were monitored. Mechanical circulatory support (MCS) was continued until five hours after the return of spontaneous circulation (ROSC), when the devices were removed. Resuscitation outcomes and the haemodynamic effects of ECPR and ICPR were compared. The cannulation time for ECMO (469 ± 129 s) was significantly longer than the time for Impella device implantation (153 ± 64 s, p < 0.001). ECPR facilitated ROSC in 6/6 animals, whereas ICPR facilitated ROSC in 6/8 animals (p = 0.19). Echocardiography revealed no difference in LV or right ventricular (RV) dysfunction between the ECPR- and ICPR-treated animals after resuscitation (LV-global longitudinal strain [GLS] 3 h post-ROSC: ICPR: - 16.5 ± 5.6% vs. ECPR: - 13.7 ± 5.9%, p = 0.99; RV-GLS 3 h post-ROSC: ICPR: - 15.9 ± 3.3% vs. ECPR: - 17.3 ± 10.6%, p = 0.99). MCS using VA-ECMO and the Impella device both provided effective haemodynamic support during CA and post-ROSC in this large animal model. Despite LV unloading conferring a hypothetical advantage for ICPR, no significant differences in myocardial recovery were observed.

Scientific ReportsVol. 15(1)
Abiomed (Germany) (DE), St.-Johannes-Hospital Dortmund (DE), RWTH Aachen University (DE)
RWTH Aachen University
Openalex Percentile: Top 16%
Cardiac Arrest and Resuscitation
2.25
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.