Accurate computational aberration correction of volumetric optical coherence tomography based on bi-pupil imaging formulation and simultaneous estimation of multiple aberration orders
This study investigates digital aberration correction (DAC) in optical coherence tomography (OCT), addressing challenges in high-resolution imaging due to defocus and aberrations. A DAC method based on OCT image formation theory is developed, simulating wavefront errors at pupils to calculate the system response function and obtain a deconvolution filter. The approach is applied to in vivo human eye volumetric imaging, comparing bi-pupil-based and conventional correction methods. Experiments using a 1-μm swept-source OCT system demonstrate improved image quality, with the bi-pupil-based method showing superior performance in visualizing photoreceptor mosaics and retinal capillaries. This technique offers efficient estimation and correction of volumetric data in point-scanning OCT with systematic aberrations.
Authors
- Xibo Wang (ORCID: https://orcid.org/0000-0002-4220-9246)
- Yoshiaki Yasuno (ORCID: https://orcid.org/0000-0003-1645-7948)
- Shuichi Makita (ORCID: https://orcid.org/0000-0002-6614-3640)
- Suzuyo Komeda
- Nobuhisa Tateno
Publication Details
- Published
- 2025-03-19
- DOI
- https://doi.org/10.1117/12.3041978
- Primary Topic
- Optical Coherence Tomography Applications
- Type
- article
- Field-Weighted Citation Impact
- 0.00