In vivo ultra-high-sensitive OCT angiography with massive-frame-repeat and hardware and software motion suppression

Optical coherence tomography angiography (OCTA) can visualize flow structures of skin. Its sensitivity depends on the number of frames per position and the time duration of the sequential frame acquisition (i.e., the time window). OCTA typically uses a few frames and a few-to-ten-millisecond time window. Although more frames and a longer time window may enable the visualization of finer vasculature, sample motion hampers OCTA measurement. We demonstrate ultra-high-sensitive (UHS-) OCTA of in vivo human skin using 32 frames per position and a very long (a few seconds) time window. A custom scanner with a sample-fixation tip and an image-correlation-based motion correction were used to restrain sample motion. The outer forearm of 10 subjects was measured using a 1,310-nm swept-source OCT. A 6 × 6 mm² lateral field was divided into 8 blocks. Each block was scanned by a 32-repeating raster scan. The time window was 6.35 seconds. We adopted three modalities for UHS-OCTA: the logarithmic intensity variance (LIV), the late OCT auto-correlation decay speed (OCDSl) and multiple-frame correlation-mapping OCA (cmOCA). We also computed complex-correlation-based OCTA. These three types of UHS-OCTAs were individually observed and fused into a pseudo-color UHS-OCTA. For reference, the sample was also scanned using the standard 4-frame OCTA protocol, and standard OCTA was computed. En face average projections were generated at depth slabs of 0 to 200-μm, 200 to 400-μm, 400 to 600-μm, and 600 to 800-μm from the skin surface. Subjective observation clarified that pseudo-color UHS-OCTA provides very high contrast visualization of fine vessels and revealed significantly more vessels than standard OCTA. The vessel density of LIV-based UHS-OCTA and standard OCTA was computed. The UHS-OCTA showed significantly higher vessel density than standard OCTA (p < 0.001 for all slabs).

Authors

Institutions

Publication Details

Published
2025-03-19
DOI
https://doi.org/10.1117/12.3041726
Primary Topic
Optical Coherence Tomography Applications
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

In vivo ultra-high-sensitive OCT angiography with massive-frame-repeat and hardware and software motion suppression

Carmen Bao, Rion Morishita, Yoshiaki Yasuno, Yiqiang Zhu et al.
Optical Coherence Tomography Applications
article

In vivo ultra-high-sensitive OCT angiography with massive-frame-repeat and hardware and software motion suppression

Carmen Bao, Rion Morishita, Yoshiaki Yasuno, Yiqiang Zhu, Guo Yu, Shuichi Makita, Pradipta Mukherjee, Ibrahim Abd El-Sadek
article en

Abstract

Optical coherence tomography angiography (OCTA) can visualize flow structures of skin. Its sensitivity depends on the number of frames per position and the time duration of the sequential frame acquisition (i.e., the time window). OCTA typically uses a few frames and a few-to-ten-millisecond time window. Although more frames and a longer time window may enable the visualization of finer vasculature, sample motion hampers OCTA measurement. We demonstrate ultra-high-sensitive (UHS-) OCTA of in vivo human skin using 32 frames per position and a very long (a few seconds) time window. A custom scanner with a sample-fixation tip and an image-correlation-based motion correction were used to restrain sample motion. The outer forearm of 10 subjects was measured using a 1,310-nm swept-source OCT. A 6 × 6 mm² lateral field was divided into 8 blocks. Each block was scanned by a 32-repeating raster scan. The time window was 6.35 seconds. We adopted three modalities for UHS-OCTA: the logarithmic intensity variance (LIV), the late OCT auto-correlation decay speed (OCDSl) and multiple-frame correlation-mapping OCA (cmOCA). We also computed complex-correlation-based OCTA. These three types of UHS-OCTAs were individually observed and fused into a pseudo-color UHS-OCTA. For reference, the sample was also scanned using the standard 4-frame OCTA protocol, and standard OCTA was computed. En face average projections were generated at depth slabs of 0 to 200-μm, 200 to 400-μm, 400 to 600-μm, and 600 to 800-μm from the skin surface. Subjective observation clarified that pseudo-color UHS-OCTA provides very high contrast visualization of fine vessels and revealed significantly more vessels than standard OCTA. The vessel density of LIV-based UHS-OCTA and standard OCTA was computed. The UHS-OCTA showed significantly higher vessel density than standard OCTA (p < 0.001 for all slabs).

University of Tsukuba (JP), Indian Institute of Technology Delhi (IN)
Openalex Percentile: Top 95%
Optical Coherence Tomography Applications
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.