Measurements of Small Field Output Correction Factors on a RefleXion Treatment Machine
The small field output correction factors of 6 kinds of detectors were determined for a RefleXion system, conforming to the formalism in TRS-483. All detectors except CC13 fulfil the 5% correction limit recommended by the TRS-483 for output factor measurement.