Measurements of Small Field Output Correction Factors on a RefleXion Treatment Machine

The small field output correction factors of 6 kinds of detectors were determined for a RefleXion system, conforming to the formalism in TRS-483. All detectors except CC13 fulfil the 5% correction limit recommended by the TRS-483 for output factor measurement.

Authors

Institutions

Publication Details

Journal
Precision Radiation Oncology
Published
2025-03-01
DOI
https://doi.org/10.1002/pro6.70006
Primary Topic
Advanced Radiotherapy Techniques
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Measurements of Small Field Output Correction Factors on a RefleXion Treatment Machine

Huixiao Chen, Fada Guan, Zhe Chen, Minyoung Lee et al.
Precision Radiation Oncology
Advanced Radiotherapy Techniques
article

Measurements of Small Field Output Correction Factors on a RefleXion Treatment Machine

Huixiao Chen, Fada Guan, Zhe Chen, Minyoung Lee, W. Zhong, Dae Yup Han
article en

Abstract

The small field output correction factors of 6 kinds of detectors were determined for a RefleXion system, conforming to the formalism in TRS-483. All detectors except CC13 fulfil the 5% correction limit recommended by the TRS-483 for output factor measurement.

Precision Radiation OncologyVol. 9(1)
Yale University (US)
Openalex Percentile: Top 93%
Advanced Radiotherapy Techniques
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.