Geometric and dosimetric evaluation of a commercial AI auto‐contouring tool on multiple anatomical sites in CT scans

Current radiotherapy practices rely on manual contouring of CT scans, which is time-consuming, prone to variability, and requires highly trained experts. There is a need for more efficient and consistent contouring methods. This study evaluated the performance of the Varian Ethos AI auto-contouring tool to assess its potential integration into clinical workflows. This retrospective study included 223 patients with treatment sites in the pelvis, abdomen, thorax, and head and neck regions. The Ethos AI tool generated auto-contours on each patients' pre-treatment planning CT, and 45 unique structures were included across the study cohort. Multiple measures of geometric similarity were computed, including surface Dice Similarity Coefficient (sDSC) and mean distance to agreement (MDA). Dosimetric concordance was evaluated by comparing mean dose and maximum 2 cm<sup>3</sup> dose (D<sub>2 cc</sub>) between manual and AI contours. Ethos AI demonstrated high geometric accuracy for well-defined structures like the bladder, lungs, and femoral heads. Smaller structures and those with less defined boundaries, such as optic nerves and duodenum, showed lower agreement. Over 70% of auto-contours demonstrated a sDSC > 0.8, and 74% had MDA < 2.5 mm. Geometric accuracy generally correlated with dosimetric concordance, however differences in contour definitions did result in some structures exhibiting dose deviations. The Ethos AI auto-contouring tool offers promising accuracy and reliability for many anatomical structures, supporting its use in planning workflows. Auto-contouring errors, although rare, highlight the importance of ongoing QA and expert manual oversight.

Authors

Institutions

Publication Details

Journal
Journal of Applied Clinical Medical Physics
Published
2025-03-17
DOI
https://doi.org/10.1002/acm2.70067
Citations
5
Primary Topic
Advanced Radiotherapy Techniques
Type
article
Field-Weighted Citation Impact
9.41
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Geometric and dosimetric evaluation of a commercial AI auto‐contouring tool on multiple anatomical sites in CT scans

Maegan Stewart, Patrick Horsley, Jeremy Booth, Regina Bromley et al.
5 citations
Journal of Applied Clinical Medical Physics
Advanced Radiotherapy Techniques
9.41
article

Geometric and dosimetric evaluation of a commercial AI auto‐contouring tool on multiple anatomical sites in CT scans

Maegan Stewart, Patrick Horsley, Jeremy Booth, Regina Bromley, Alexandra Quinn, Robert Finnegan, Joseph Chan
article en
5 citations

Abstract

Current radiotherapy practices rely on manual contouring of CT scans, which is time-consuming, prone to variability, and requires highly trained experts. There is a need for more efficient and consistent contouring methods. This study evaluated the performance of the Varian Ethos AI auto-contouring tool to assess its potential integration into clinical workflows. This retrospective study included 223 patients with treatment sites in the pelvis, abdomen, thorax, and head and neck regions. The Ethos AI tool generated auto-contours on each patients' pre-treatment planning CT, and 45 unique structures were included across the study cohort. Multiple measures of geometric similarity were computed, including surface Dice Similarity Coefficient (sDSC) and mean distance to agreement (MDA). Dosimetric concordance was evaluated by comparing mean dose and maximum 2 cm<sup>3</sup> dose (D<sub>2 cc</sub>) between manual and AI contours. Ethos AI demonstrated high geometric accuracy for well-defined structures like the bladder, lungs, and femoral heads. Smaller structures and those with less defined boundaries, such as optic nerves and duodenum, showed lower agreement. Over 70% of auto-contours demonstrated a sDSC > 0.8, and 74% had MDA < 2.5 mm. Geometric accuracy generally correlated with dosimetric concordance, however differences in contour definitions did result in some structures exhibiting dose deviations. The Ethos AI auto-contouring tool offers promising accuracy and reliability for many anatomical structures, supporting its use in planning workflows. Auto-contouring errors, although rare, highlight the importance of ongoing QA and expert manual oversight.

Journal of Applied Clinical Medical PhysicsVol. 26(6)
The University of Sydney (AU), Royal North Shore Hospital (AU)
Openalex Percentile: Top 2%
Advanced Radiotherapy Techniques
9.41
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.