The model to correct for the ion recombination of micro-thin cylindrical ionization chamber in high dose-rate and frequency pulses X-ray beams
Accurate dosimetry is essential for ultrahigh dose-rate (UHDR, >40 Gy/s) irradiations in clinical radiotherapy (RT). Herein, we introduced a prototype of a micro-thin cylindrical ionization chamber (MC-IC) featuring an electrode distance of 0.2 mm, specifically designed for these beams. This study aimed to investigate the ion collection efficiency (ICE) of the MC-IC in high dose-rate and frequency (HDR&F) pulse X-ray beams. A detailed comparison was made among an empirical logistic model, the two-voltage analysis (TVA) method, and the Boag model. As anticipated, the MC-IC successfully measured a mean dose rate of up to 154.5 Gy/s with an ICE of 98.6 %. We found that the empirical logistic model is well-suited for the MC-IC, whereas the TVA method and Boag model exhibit an underestimation of ICE in HDR&F beams. Therefore, Boag Model 3 with a free-electron fraction of 0.5 < P < 0.6 and a modified TVA equation with a free-electron fraction of P = 0.6 must be considered. For the TVA method, a voltage ratio ≤2 is recommended. In conclusion, this study complemented and refined a dosimetric methodology for evaluating the collection efficiency correction factor for special ionization chambers. • A MC-IC designed for UDR&F X-ray beams, achieving a mean dose rate of up to 154.5 Gy/s with an ICE of 98.6%. • We assessed three ion recombination models for MC-IC and the empirical logistic model is suitable for MI-IC. • A voltage ratio ≤ 2 and Equation (9) with p = 0.6 Model 3is recommended for the TVA mothed in HDR&F X-ray beams. • A free-electron fraction of 0.5 < p < 0.6 must be considered for the Boag Model 3 in HDR&F X-ray beams.
Authors
- Sen Bai (ORCID: https://orcid.org/0000-0002-7146-6970)
- Wanchang LAI
- Yanmei HAO
Institutions
- University of Electronic Science and Technology of China (CN)
- Sichuan University (CN)
- Chengdu University of Technology (CN)
- West China Hospital of Sichuan University (CN)
- Sichuan Cancer Hospital (CN)
Publication Details
- Journal
- Nuclear Instruments and Methods in Physics Research Section A Accelerators Spectrometers Detectors and Associated Equipment
- Published
- 2025-03-17
- DOI
- https://doi.org/10.1016/j.nima.2025.170434
- Primary Topic
- Radiation Therapy and Dosimetry
- Type
- article
- Field-Weighted Citation Impact
- 0.00