Robot milling system integrated design and finite element analysis of custom femoral prostheses

The long-term stability of cementless femoral prostheses is primarily affected by aseptic loosening, micromotion, and stress shielding, all of which are related to the force transfer of the prosthesis. These factors can compromise the osseointegration of the proximal prosthesis, leading to aseptic loosening within the femoral cavity. Due to the individual variability of the femur, the fit between the prosthesis and the femoral cavity during the design phase may differ from the fit achieved during the surgical procedure. Consequently, the force transfer of the prosthesis postoperatively may not align with the results obtained from finite element analysis conducted during the design phase, making it challenging to control the micromotion and stress shielding of the prosthesis. The design model of a custom femoral prosthesis is based on the CT reconstruction of the patient' femur. The fit of prosthesis within the femoral cavity during the design phase should match the fit during the surgical operation. Consequently, the results of finite element analysis conducted during the design phase can be used to control the force transfer of the prosthesis postoperatively. This approach helps to prevent improper micromotion and stress shielding of the proximal prosthesis, which can compromise the primary stability of the prosthesis within the femoral cavity, thereby facilitating the osseointegration of the proximal prosthesis.This paper proposes a novel technology that combines the design, finite element analysis, and manufacturing of custom prostheses. Specifically, a CAD/CAM/robot integration method is used to fabricate these prostheses. This innovative technology not only enhances the control of force transfer in custom prostheses but also reduces design and manufacture time while lowering costs. In conclusion, the finite element analysis of the custom prosthesis effectively manages force transfer, and the milling errors associated with the custom prosthesis are less than 1 mm.

Authors

Institutions

Publication Details

Journal
Journal of Orthopaedic Surgery and Research
Published
2025-03-18
DOI
https://doi.org/10.1186/s13018-025-05647-w
Primary Topic
Orthopaedic implants and arthroplasty
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Robot milling system integrated design and finite element analysis of custom femoral prostheses

Renjie He, Si‐Yu Yin, Qilin Wu, Wang Liu et al.
Journal of Orthopaedic Surgery and Research
Orthopaedic implants and arthroplasty
article

Robot milling system integrated design and finite element analysis of custom femoral prostheses

Renjie He, Si‐Yu Yin, Qilin Wu, Wang Liu, Zhenjie Li, Pengju Yue
article en

Abstract

The long-term stability of cementless femoral prostheses is primarily affected by aseptic loosening, micromotion, and stress shielding, all of which are related to the force transfer of the prosthesis. These factors can compromise the osseointegration of the proximal prosthesis, leading to aseptic loosening within the femoral cavity. Due to the individual variability of the femur, the fit between the prosthesis and the femoral cavity during the design phase may differ from the fit achieved during the surgical procedure. Consequently, the force transfer of the prosthesis postoperatively may not align with the results obtained from finite element analysis conducted during the design phase, making it challenging to control the micromotion and stress shielding of the prosthesis. The design model of a custom femoral prosthesis is based on the CT reconstruction of the patient' femur. The fit of prosthesis within the femoral cavity during the design phase should match the fit during the surgical operation. Consequently, the results of finite element analysis conducted during the design phase can be used to control the force transfer of the prosthesis postoperatively. This approach helps to prevent improper micromotion and stress shielding of the proximal prosthesis, which can compromise the primary stability of the prosthesis within the femoral cavity, thereby facilitating the osseointegration of the proximal prosthesis.This paper proposes a novel technology that combines the design, finite element analysis, and manufacturing of custom prostheses. Specifically, a CAD/CAM/robot integration method is used to fabricate these prostheses. This innovative technology not only enhances the control of force transfer in custom prostheses but also reduces design and manufacture time while lowering costs. In conclusion, the finite element analysis of the custom prosthesis effectively manages force transfer, and the milling errors associated with the custom prosthesis are less than 1 mm.

Journal of Orthopaedic Surgery and ResearchVol. 20(1)
82th Hospital of Pla (CN)
Openalex Percentile: Top 91%
Orthopaedic implants and arthroplasty
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.