Accuracy of Two Continuous Glucose Monitors Differs after Hydroxyurea in Pediatric Patients Undergoing Total Pancreatectomy with Islet Autotransplantation

<b><i>Background:</i></b> Total pancreatectomy with islet autotransplantation (TPIAT) requires strict glycemic management for islet survival using insulin pumps and continuous glucose monitors (CGMs). Hydroxyurea prevents reactive thrombocytosis but interferes with the accuracy of the Dexcom CGM. Hydroxyurea is reported to not interfere with the Libre CGM but has not been studied after TPIAT. <b><i>Methods:</i></b> Seven patients wore both Dexcom and Libre starting approximately a week after TPIAT. Dexcom and Libre values were obtained with point-of-care testing blood glucose (POCT BG) at 560 unique time points. Descriptive statistics included median, interquartile range (IQR), absolute difference between CGM and POCT, and mean absolute relative difference (MARD) for each Dexcom and Libre. Wilcoxon-Mann-Whitney tests were performed to compare parameters between Dexcom and Libre, with two-sided significance of <i>P</i> < 0.05. Clarke error grids and boxplots were constructed. <b><i>Results:</i></b> In the 9 h after hydroxyurea, median POCT BG was 110 mg/dL (IQR 88-143), median Dexcom BG was 172 mg/dL (135-219), and median Libre BG was 106 mg/dL (76-138). MARD for Dexcom was 59.5% and for Libre was 14.8% (<i>P</i> < 0.001). Median absolute difference between Dexcom and POCT BG (56 mg/dL [32-88]) was greater than that for Libre (12 mg/dL [6-23]; <i>P</i> < 0.001). In Clarke error grids, 98.3% of values fell within clinically acceptable Zones A/B for Libre; 77.9% of values fell within these zones for Dexcom. At all other times, median POCT BG was 110 mg/dL (86-133), median Dexcom BG was 124 mg/dL (97-154), and median Libre BG was 104 mg/dL (76-128). MARD for Dexcom was 19.8% and for Libre was 14.7% (<i>P</i> < 0.001). Median absolute difference between Dexcom and POCT BG (18 mg/dL [9-30]) was clinically similar to that for Libre (13 mg/dL [6-23], <i>P</i> < 0.001). <b><i>Conclusion:</i></b> Hydroxyurea does not seem to interfere with the accuracy of Libre in contrast to Dexcom. Use of Libre after TPIAT could facilitate improved glycemic management.

Authors

Institutions

Publication Details

Journal
Diabetes Technology & Therapeutics
Published
2025-03-18
DOI
https://doi.org/10.1089/dia.2025.0010
Primary Topic
Pancreatic function and diabetes
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Accuracy of Two Continuous Glucose Monitors Differs after Hydroxyurea in Pediatric Patients Undergoing Total Pancreatectomy with Islet Autotransplantation

Cheryl E. Gariepy, Kajal Gandhi, Sara K. Rasmussen, Christie Heinzman et al.
2 links
Diabetes Technology & Therapeutics
Pancreatic function and diabetes
article

Accuracy of Two Continuous Glucose Monitors Differs after Hydroxyurea in Pediatric Patients Undergoing Total Pancreatectomy with Islet Autotransplantation

Cheryl E. Gariepy, Kajal Gandhi, Sara K. Rasmussen, Christie Heinzman, Robert P. Hoffman, Jaimie D. Nathan, Jennifer M. Ladd, A. Jay Freeman, Mary Joy Okafor, Kelly Friesner-Gephart
article en

Abstract

<b><i>Background:</i></b> Total pancreatectomy with islet autotransplantation (TPIAT) requires strict glycemic management for islet survival using insulin pumps and continuous glucose monitors (CGMs). Hydroxyurea prevents reactive thrombocytosis but interferes with the accuracy of the Dexcom CGM. Hydroxyurea is reported to not interfere with the Libre CGM but has not been studied after TPIAT. <b><i>Methods:</i></b> Seven patients wore both Dexcom and Libre starting approximately a week after TPIAT. Dexcom and Libre values were obtained with point-of-care testing blood glucose (POCT BG) at 560 unique time points. Descriptive statistics included median, interquartile range (IQR), absolute difference between CGM and POCT, and mean absolute relative difference (MARD) for each Dexcom and Libre. Wilcoxon-Mann-Whitney tests were performed to compare parameters between Dexcom and Libre, with two-sided significance of <i>P</i> < 0.05. Clarke error grids and boxplots were constructed. <b><i>Results:</i></b> In the 9 h after hydroxyurea, median POCT BG was 110 mg/dL (IQR 88-143), median Dexcom BG was 172 mg/dL (135-219), and median Libre BG was 106 mg/dL (76-138). MARD for Dexcom was 59.5% and for Libre was 14.8% (<i>P</i> < 0.001). Median absolute difference between Dexcom and POCT BG (56 mg/dL [32-88]) was greater than that for Libre (12 mg/dL [6-23]; <i>P</i> < 0.001). In Clarke error grids, 98.3% of values fell within clinically acceptable Zones A/B for Libre; 77.9% of values fell within these zones for Dexcom. At all other times, median POCT BG was 110 mg/dL (86-133), median Dexcom BG was 124 mg/dL (97-154), and median Libre BG was 104 mg/dL (76-128). MARD for Dexcom was 19.8% and for Libre was 14.7% (<i>P</i> < 0.001). Median absolute difference between Dexcom and POCT BG (18 mg/dL [9-30]) was clinically similar to that for Libre (13 mg/dL [6-23], <i>P</i> < 0.001). <b><i>Conclusion:</i></b> Hydroxyurea does not seem to interfere with the accuracy of Libre in contrast to Dexcom. Use of Libre after TPIAT could facilitate improved glycemic management.

Diabetes Technology & TherapeuticsVol. 27(8)
Nationwide Children's Hospital (US), The Ohio State University (US)
Zero hunger
Openalex Percentile: Top 91%
Pancreatic function and diabetes
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.