Implant Body Fracture Due to Fatigue in a Maxillary Implant‐Retained Overdenture Treatment

The aim of this study was to determine the causes of fatigue and implant failures in an implant-retained overdenture case. One out of four implants portrayed inadequate osseointegration during healing abutment placement, and one was fractured from the middle to the apical third after 1 year of denture use. Various aspects, such as insufficient cooling while implant socket preparation, incorrect prosthodontic planning, and erroneous design of the denture, were evaluated. Macroscopic and microscopic inspections of the fractured implant body were assessed by scanning electron microscopy (SEM) and optical microscopy. Consequently, metallic fatigue along with secondary cracks and beach marks were evident. It was deduced that the failed and fractured implants depicted substandard osseointegration; thus, they were susceptible to the wear out effect of the occlusal forces. Unfavorable osseointegration caused the fulcrum axis to move apically instead of forming on the neck part of the implant, and micromovements around the fulcrum axis mustered the fracture.

Authors

Institutions

Publication Details

Journal
Case Reports in Dentistry
Published
2025-01-01
DOI
https://doi.org/10.1155/crid/4064452
Citations
1
Primary Topic
Dental Implant Techniques and Outcomes
Type
article
Field-Weighted Citation Impact
1.79
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Implant Body Fracture Due to Fatigue in a Maxillary Implant‐Retained Overdenture Treatment

Hakan Terzioğlu, Lale Karaağaçlıoğlu, Rıza Gürbüz, Can Hakan Sarıkaya
1 citations
Case Reports in Dentistry
Dental Implant Techniques and Outcomes
1.79
article

Implant Body Fracture Due to Fatigue in a Maxillary Implant‐Retained Overdenture Treatment

Hakan Terzioğlu, Lale Karaağaçlıoğlu, Rıza Gürbüz, Can Hakan Sarıkaya
article en
1 citations

Abstract

The aim of this study was to determine the causes of fatigue and implant failures in an implant-retained overdenture case. One out of four implants portrayed inadequate osseointegration during healing abutment placement, and one was fractured from the middle to the apical third after 1 year of denture use. Various aspects, such as insufficient cooling while implant socket preparation, incorrect prosthodontic planning, and erroneous design of the denture, were evaluated. Macroscopic and microscopic inspections of the fractured implant body were assessed by scanning electron microscopy (SEM) and optical microscopy. Consequently, metallic fatigue along with secondary cracks and beach marks were evident. It was deduced that the failed and fractured implants depicted substandard osseointegration; thus, they were susceptible to the wear out effect of the occlusal forces. Unfavorable osseointegration caused the fulcrum axis to move apically instead of forming on the neck part of the implant, and micromovements around the fulcrum axis mustered the fracture.

Case Reports in DentistryVol. 2025(1)
Ankara University (TR), Metallurgical Research Institute (RO), Lokman Hekim Üniversitesi (TR)
Life below water
Openalex Percentile: Top 19%
Dental Implant Techniques and Outcomes
1.79
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.