A Method to Calibrate Angular Positioning Errors Using a Laser Tracker and a Plane Mirror

We describe a method to calibrate angular positioning errors of a rotation stage using a laser tracker (LT), a plane mirror mounted on the stage, and stationary registration nests placed around the stage. Our technique involves determining the direction of the normal vector to the plane of the mirror at each angular step by performing two LT measurements-one directly to a stationary spherically mounted retroreflector (SMR), and another to the same SMR by bouncing the laser off a mirror mounted on the rotation stage. Because the angular range that can be measured from a single LT station is limited by the angle of incidence on the mirror, multiple LT stations are necessary to cover the full 360°, hence the need for stationary registration nests to tie the LT data into a common coordinate frame. We compare this technique against a direct approach involving a rigid bar with two SMRs mounted on the rotation stage so that we can measure the direction of the line joining the SMRs at each angular position using the LT and, therefore, the angle between positions. Through experiments, we demonstrate that our mirror-based approach provides errors on the order of ±0.5″, smaller than the ±1.5″ for the direct approach, when compared against a reference instrument with accuracy better than 0.3″. Through simulations, we estimate the uncertainty in our mirror-based angle measurements to be 0.4″ (<i>k</i> = 2). Placing the LT close to and the SMR away from the rotation stage results in lower uncertainty for our mirror-based angle measurements.

Authors

Institutions

Publication Details

Journal
Sensors
Published
2025-03-15
DOI
https://doi.org/10.3390/s25061834
Citations
2
Primary Topic
Advanced Measurement and Metrology Techniques
Type
article
Field-Weighted Citation Impact
1.43
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

A Method to Calibrate Angular Positioning Errors Using a Laser Tracker and a Plane Mirror

Bala Muralikrishnan, Glenn Holland, John S. Villarrubia, Olga Ridzel et al.
2 citations
Sensors
Advanced Measurement and Metrology Techniques
1.43
article

A Method to Calibrate Angular Positioning Errors Using a Laser Tracker and a Plane Mirror

Bala Muralikrishnan, Glenn Holland, John S. Villarrubia, Olga Ridzel, Vincent Lee, Meghan Shilling
article en
2 citations

Abstract

We describe a method to calibrate angular positioning errors of a rotation stage using a laser tracker (LT), a plane mirror mounted on the stage, and stationary registration nests placed around the stage. Our technique involves determining the direction of the normal vector to the plane of the mirror at each angular step by performing two LT measurements-one directly to a stationary spherically mounted retroreflector (SMR), and another to the same SMR by bouncing the laser off a mirror mounted on the rotation stage. Because the angular range that can be measured from a single LT station is limited by the angle of incidence on the mirror, multiple LT stations are necessary to cover the full 360°, hence the need for stationary registration nests to tie the LT data into a common coordinate frame. We compare this technique against a direct approach involving a rigid bar with two SMRs mounted on the rotation stage so that we can measure the direction of the line joining the SMRs at each angular position using the LT and, therefore, the angle between positions. Through experiments, we demonstrate that our mirror-based approach provides errors on the order of ±0.5″, smaller than the ±1.5″ for the direct approach, when compared against a reference instrument with accuracy better than 0.3″. Through simulations, we estimate the uncertainty in our mirror-based angle measurements to be 0.4″ (<i>k</i> = 2). Placing the LT close to and the SMR away from the rotation stage results in lower uncertainty for our mirror-based angle measurements.

SensorsVol. 25(6)
National Institute of Standards and Technology (US), Theiss Research (US)
Openalex Percentile: Top 22%
Advanced Measurement and Metrology Techniques
1.43
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

A Method to Calibrate Angular Positioning Errors Using a Laser Tracker and a Plane Mirror — Bala Muralikrishnan, Glenn Holland, et al. · Sensors (2025) | TGRS Research Map | TGRS