Optimizing Radiation Dose in High-Resolution Chest CT: The Impact of Patient-Specific Factors and Size-Specific Dose Estimates

<b>Background/Objectives:</b> High-resolution chest computed tomography (HRCT) is a critical diagnostic tool, but radiation dose optimization remains a significant concern. Traditional dose metrics such as the volume CT dose index (CTDIvol) and dose-length product (DLP) do not adequately account for patient size variations. This study aimed to assess the radiation dose in HRCT using size-specific dose estimates (SSDEs) and evaluate the influence of patient-specific factors on key dosimetric parameters. <b>Methods</b>: This retrospective cohort study analyzed HRCT scans from 1970 adult patients conducted between September 2022 and February 2024. Radiation dose data, including the CTDIvol, DLP, SSDE, and effective dose, were extracted from the DoseWatch™ software. Patient demographics, scan protocols, and exposure parameters were collected. Descriptive statistics, correlation analyses, and significance testing were conducted using IBM SPSS (Version 26). <b>Results</b>: A significant positive correlation was found between the radiation dose parameters (CTDIvol, DLP, SSDE) and patient body size metrics, particularly BMI (rpb = 0.445, <i>p</i> < 0.01). The SSDE values ranged from 2.7 to 12.4 mGy, providing a more patient-specific dose assessment than traditional indices. Gender differences were observed, with male patients receiving higher radiation doses (<i>p</i> < 0.01). The scanning range exhibited a significant negative correlation with the CTDIvol and SSDE, suggesting dose variations with anatomical coverage. <b>Conclusions</b>: SSDEs provide a more accurate, patient-centered dose assessment in HRCT, allowing for optimized radiation safety strategies. These findings emphasize the need for size-adapted scan protocols to minimize exposure while maintaining diagnostic image quality. The routine integration of SSDE into clinical practice is recommended to enhance individualized dose management in HRCT.

Authors

Institutions

Publication Details

Journal
Diagnostics
Published
2025-03-16
DOI
https://doi.org/10.3390/diagnostics15060740
Citations
7
Primary Topic
Radiation Dose and Imaging
Type
article
Field-Weighted Citation Impact
9.52
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Optimizing Radiation Dose in High-Resolution Chest CT: The Impact of Patient-Specific Factors and Size-Specific Dose Estimates

Mohamed Abuzaid
7 citations
Diagnostics
Radiation Dose and Imaging
9.52
article

Optimizing Radiation Dose in High-Resolution Chest CT: The Impact of Patient-Specific Factors and Size-Specific Dose Estimates

Mohamed Abuzaid
article en
7 citations

Abstract

<b>Background/Objectives:</b> High-resolution chest computed tomography (HRCT) is a critical diagnostic tool, but radiation dose optimization remains a significant concern. Traditional dose metrics such as the volume CT dose index (CTDIvol) and dose-length product (DLP) do not adequately account for patient size variations. This study aimed to assess the radiation dose in HRCT using size-specific dose estimates (SSDEs) and evaluate the influence of patient-specific factors on key dosimetric parameters. <b>Methods</b>: This retrospective cohort study analyzed HRCT scans from 1970 adult patients conducted between September 2022 and February 2024. Radiation dose data, including the CTDIvol, DLP, SSDE, and effective dose, were extracted from the DoseWatch™ software. Patient demographics, scan protocols, and exposure parameters were collected. Descriptive statistics, correlation analyses, and significance testing were conducted using IBM SPSS (Version 26). <b>Results</b>: A significant positive correlation was found between the radiation dose parameters (CTDIvol, DLP, SSDE) and patient body size metrics, particularly BMI (rpb = 0.445, <i>p</i> < 0.01). The SSDE values ranged from 2.7 to 12.4 mGy, providing a more patient-specific dose assessment than traditional indices. Gender differences were observed, with male patients receiving higher radiation doses (<i>p</i> < 0.01). The scanning range exhibited a significant negative correlation with the CTDIvol and SSDE, suggesting dose variations with anatomical coverage. <b>Conclusions</b>: SSDEs provide a more accurate, patient-centered dose assessment in HRCT, allowing for optimized radiation safety strategies. These findings emphasize the need for size-adapted scan protocols to minimize exposure while maintaining diagnostic image quality. The routine integration of SSDE into clinical practice is recommended to enhance individualized dose management in HRCT.

DiagnosticsVol. 15(6)
University of Sharjah (AE)
Good health and well-being
Openalex Percentile: Top 2%
Radiation Dose and Imaging
9.52
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.