Stand for Certification of X-ray Optical Elements and Systems for Synchrotron Applications

The article describes an X-ray optical certification setup based on a microfocus X-ray tube, a Kirkpatrick-Baez (KB) collimating system, and a Gpixel2020BSI matrix semiconductor detector. To minimize X-ray beam scattering and intensity loss due to absorption in air, vacuum-tight tubes with compton windows are installed between the microfocus X-ray tube and the KB system, and between the KB system and the detector. Pumping is performed by an oil-free fore-vacuum pump. The setup that has been developed allows for studying the X-ray optical scheme of two-mirror monochromators to determine the deformation of the beam wave front at their outputs, and to calibrate the movement of high-precision mechanical tables. The setup also allows for studying the focusing properties of the Kirkpatrick-Baez system, itself, namely, measuring the size of the focus spot and calibrating the piezo-drives of the mirrors. As a result of adjusting the collimating system, we obtained a converging beam, but this did not prevent us from measuring the wave front broadening upon reflection from the substrates for a two-mirror monochromator. Two substrates made of single-crystal silicon Si(100) were studied. It was shown that upon reflection from the substrate after chemical-mechanical polishing (CMP) only, the reflected beam, due to mirror shape errors, broadened by 18 μm, while after chemical-mechanical polishing and correction of local shape errors using an ion beam, the broadening was reduced to 11 μm, which corresponds to an angular error of the mirror at a level of 3 μrad. This is 7 times less than the angular divergence of the beam at the output of the collimating system, due to the size of the radiation source.

Authors

Institutions

Publication Details

Journal
Journal of Surface Investigation X-ray Synchrotron and Neutron Techniques
Published
2024-12-01
DOI
https://doi.org/10.1134/s1027451024701842
Primary Topic
Advanced X-ray Imaging Techniques
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Stand for Certification of X-ray Optical Elements and Systems for Synchrotron Applications

A. K. Chernyshev, E. V. Petrakov, И. В. Малышев, Egor I. Glushkov et al.
2 links
Journal of Surface Investigation X-ray Synchrotron and Neutron Techniques
Advanced X-ray Imaging Techniques
article

Stand for Certification of X-ray Optical Elements and Systems for Synchrotron Applications

A. K. Chernyshev, E. V. Petrakov, И. В. Малышев, Egor I. Glushkov, D. G. Reunov, Н. И. Чхало, I. G. Zabrodin, M. S. Mikhailenko, A. D. Akhsakhalyan
article en

Abstract

The article describes an X-ray optical certification setup based on a microfocus X-ray tube, a Kirkpatrick-Baez (KB) collimating system, and a Gpixel2020BSI matrix semiconductor detector. To minimize X-ray beam scattering and intensity loss due to absorption in air, vacuum-tight tubes with compton windows are installed between the microfocus X-ray tube and the KB system, and between the KB system and the detector. Pumping is performed by an oil-free fore-vacuum pump. The setup that has been developed allows for studying the X-ray optical scheme of two-mirror monochromators to determine the deformation of the beam wave front at their outputs, and to calibrate the movement of high-precision mechanical tables. The setup also allows for studying the focusing properties of the Kirkpatrick-Baez system, itself, namely, measuring the size of the focus spot and calibrating the piezo-drives of the mirrors. As a result of adjusting the collimating system, we obtained a converging beam, but this did not prevent us from measuring the wave front broadening upon reflection from the substrates for a two-mirror monochromator. Two substrates made of single-crystal silicon Si(100) were studied. It was shown that upon reflection from the substrate after chemical-mechanical polishing (CMP) only, the reflected beam, due to mirror shape errors, broadened by 18 μm, while after chemical-mechanical polishing and correction of local shape errors using an ion beam, the broadening was reduced to 11 μm, which corresponds to an angular error of the mirror at a level of 3 μrad. This is 7 times less than the angular divergence of the beam at the output of the collimating system, due to the size of the radiation source.

Journal of Surface Investigation X-ray Synchrotron and Neutron TechniquesVol. 18(S1)
Institute for Physics of Microstructures (RU)
Openalex Percentile: Top 64%
Advanced X-ray Imaging Techniques
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.