SIMULTANEOUS DETERMINATION OF GLYCERINE AND DEG IN GLYCERINE RAW MATERIAL USING FT-IR SPECTROSCOPY AND MULTIVARIATE CALIBRATION METHODS

Objective: This study aims to determine reliably the concentration of Diethylene Glycol (DEG) in glycerine raw material in a simple and rapid manner using Attenuated Total Reflectance-Fourier Transform Infrared (ATR-FTIR) spectroscopy coupled with chemometric methods. Methods: for the first time, the current work employs multivariate data analysis tools and FTIR spectroscopy for the simultaneous determination of glycerine and DEG. Binary mixtures of the two substances were prepared and categorized into calibration, validation and test samples. The DEG ratio in these samples ranged from 1 to 30%. The FTIR spectra were recorded for all samples in the range 3700-815 cm-1, and spectral data for each group of samples were used for Principal Component Analysis (PCA), Principal Component Regression (PCR) and Partial Least Square (PLS) applications. Results: In the case of PCA application, a 100% successful discrimination among pure glycerine, DEG and mixed samples was achieved. Semi-quantitative determination of both substances was also possible using PC1 and PC2, accounting for about 99% of the variation in the data set. When using the PCR algorithm, quantitative analysis of the two substances was successfully achieved with an average recovery percent of 98.32±0.19% for glycerine and 99.89±5.7% for DEG. On the other hand, the mean recovery percent values based on the PLS model were 101.37±0.13% and 103.26±3.2% for glycerine and DEG, respectively. Conclusion: Distinguishing pure samples of both analytes was achieved successfully using FTIR spectroscopy and PCA for data analysis. On the other hand, the quantification of DEG in glycerine was very satisfactory upon analyzing the spectroscopic data using PCR and PLS algorithms. FTIR spectroscopy coupled with PCA, PCR and PLS has, thus, been shown to be of great potential to detect the adulteration of glycerine with DEG.

Authors

Institutions

Publication Details

Journal
International Journal of Applied Pharmaceutics
Published
2025-03-07
DOI
https://doi.org/10.22159/ijap.2025v17i2.52729
Primary Topic
Advanced Chemical Sensor Technologies
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

SIMULTANEOUS DETERMINATION OF GLYCERINE AND DEG IN GLYCERINE RAW MATERIAL USING FT-IR SPECTROSCOPY AND MULTIVARIATE CALIBRATION METHODS

Safwan M. Obeidat, Ayman Hammoudeh
International Journal of Applied Pharmaceutics
Advanced Chemical Sensor Technologies
article

SIMULTANEOUS DETERMINATION OF GLYCERINE AND DEG IN GLYCERINE RAW MATERIAL USING FT-IR SPECTROSCOPY AND MULTIVARIATE CALIBRATION METHODS

Safwan M. Obeidat, Ayman Hammoudeh
article en

Abstract

Objective: This study aims to determine reliably the concentration of Diethylene Glycol (DEG) in glycerine raw material in a simple and rapid manner using Attenuated Total Reflectance-Fourier Transform Infrared (ATR-FTIR) spectroscopy coupled with chemometric methods. Methods: for the first time, the current work employs multivariate data analysis tools and FTIR spectroscopy for the simultaneous determination of glycerine and DEG. Binary mixtures of the two substances were prepared and categorized into calibration, validation and test samples. The DEG ratio in these samples ranged from 1 to 30%. The FTIR spectra were recorded for all samples in the range 3700-815 cm-1, and spectral data for each group of samples were used for Principal Component Analysis (PCA), Principal Component Regression (PCR) and Partial Least Square (PLS) applications. Results: In the case of PCA application, a 100% successful discrimination among pure glycerine, DEG and mixed samples was achieved. Semi-quantitative determination of both substances was also possible using PC1 and PC2, accounting for about 99% of the variation in the data set. When using the PCR algorithm, quantitative analysis of the two substances was successfully achieved with an average recovery percent of 98.32±0.19% for glycerine and 99.89±5.7% for DEG. On the other hand, the mean recovery percent values based on the PLS model were 101.37±0.13% and 103.26±3.2% for glycerine and DEG, respectively. Conclusion: Distinguishing pure samples of both analytes was achieved successfully using FTIR spectroscopy and PCA for data analysis. On the other hand, the quantification of DEG in glycerine was very satisfactory upon analyzing the spectroscopic data using PCR and PLS algorithms. FTIR spectroscopy coupled with PCA, PCR and PLS has, thus, been shown to be of great potential to detect the adulteration of glycerine with DEG.

International Journal of Applied Pharmaceutics
Yarmouk University (JO)
Openalex Percentile: Top 95%
Advanced Chemical Sensor Technologies
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.