<b>Background</b>: The increasing detection of non-palpable breast lesions necessitates accurate preoperative localization to ensure complete excision while preserving healthy tissue and optimizing cosmetic outcomes. Traditional wire-guided localization (WL) has been the gold standard; however, it has several drawbacks, including patient discomfort and scheduling challenges. This study evaluates the accuracy and feasibility of radiofrequency identification (RFID) tag localization using the Hologic LOCalizer™ system as an alternative technique. <b>Methods</b>: This retrospective study included 258 consecutive patients who underwent image-guided RFID tag localization from March 2021 to February 2023 from a single-center London breast unit. The primary outcome measured was the accuracy of RFID tag placement, defined as within 10 mm of the target lesion on post-clip mammograms. Secondary outcomes included type and size of lesions, re-excision rates, review of post-operative specimen radiographs, and patient demographics. <b>Results</b>: A total of 297 RFID tags were placed, with 95.6% accurately positioned within the target range. The median target size was 29 mm, with the most common abnormalities being mass lesions (64%). Among the 13 inaccurately placed RFID tags (4.4%), all were identified preoperatively, with two requiring additional wire placements. RFID tags were successfully identified in 92% of specimen radiographs, and 8% of patients required re-excision due to positive or close margins. Notably, patients with multiple RFID tags showed a higher incidence of re-excision. <b>Conclusions</b>: The LOCalizer™ RFID system demonstrated a high accuracy rate for preoperative localization of breast lesions, presenting a viable alternative to WL. This technique improves surgical scheduling flexibility and enhances patient comfort. Comparative studies with other wire-free localization technologies, such as magnetic seeds and radar reflectors, are needed to determine the optimal approach for clinical practice.

Authors

Institutions

Publication Details

Journal
Diagnostics
Published
2025-03-17
DOI
https://doi.org/10.3390/diagnostics15060746
Citations
4
Primary Topic
Breast Cancer Treatment Studies
Type
article
Field-Weighted Citation Impact
3.75
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Evaluation of Hologic LOCalizer™ RFID Tags for Preoperative Localization of Breast Lesions: A Single-Center Experience

Kefah Mokbel, Sawsan Taif, Arvind Menon, Charlotte Munday et al.
4 citations
Diagnostics
Breast Cancer Treatment Studies
3.75
article

Evaluation of Hologic LOCalizer™ RFID Tags for Preoperative Localization of Breast Lesions: A Single-Center Experience

Kefah Mokbel, Sawsan Taif, Arvind Menon, Charlotte Munday, Adeola Omotade, Anmol Malhotra
article en
4 citations

Abstract

<b>Background</b>: The increasing detection of non-palpable breast lesions necessitates accurate preoperative localization to ensure complete excision while preserving healthy tissue and optimizing cosmetic outcomes. Traditional wire-guided localization (WL) has been the gold standard; however, it has several drawbacks, including patient discomfort and scheduling challenges. This study evaluates the accuracy and feasibility of radiofrequency identification (RFID) tag localization using the Hologic LOCalizer™ system as an alternative technique. <b>Methods</b>: This retrospective study included 258 consecutive patients who underwent image-guided RFID tag localization from March 2021 to February 2023 from a single-center London breast unit. The primary outcome measured was the accuracy of RFID tag placement, defined as within 10 mm of the target lesion on post-clip mammograms. Secondary outcomes included type and size of lesions, re-excision rates, review of post-operative specimen radiographs, and patient demographics. <b>Results</b>: A total of 297 RFID tags were placed, with 95.6% accurately positioned within the target range. The median target size was 29 mm, with the most common abnormalities being mass lesions (64%). Among the 13 inaccurately placed RFID tags (4.4%), all were identified preoperatively, with two requiring additional wire placements. RFID tags were successfully identified in 92% of specimen radiographs, and 8% of patients required re-excision due to positive or close margins. Notably, patients with multiple RFID tags showed a higher incidence of re-excision. <b>Conclusions</b>: The LOCalizer™ RFID system demonstrated a high accuracy rate for preoperative localization of breast lesions, presenting a viable alternative to WL. This technique improves surgical scheduling flexibility and enhances patient comfort. Comparative studies with other wire-free localization technologies, such as magnetic seeds and radar reflectors, are needed to determine the optimal approach for clinical practice.

DiagnosticsVol. 15(6)
Royal Free London NHS Foundation Trust (GB), The Royal Free Hospital (GB), Breast Cancer Care (GB), Whittington Health NHS Trust (GB), University College London (GB)
Openalex Percentile: Top 7%
Breast Cancer Treatment Studies
3.75
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.