Fracture of the Femoral Broach Locking Mechanism and Surgical Technique to Retrieve the Broach

The current case series reports on 4 cases of intraoperative fracture of the femoral broach locking mechanism during total hip arthroplasty and subsequent entrapment of the broach. Three of the cases were performed via the direct anterior approach and 1 was performed with the posterolateral approach. Two femoral broach locking mechanisms fractured with the use of a power broaching device and 2 occurred when using the calcar planar. The authors also discuss a minimally invasive technique to remove the broach. While intraoperative fracture of the femoral broach locking mechanism is not a common complication, surgeons should be aware of the potential causes of these complications as well as strategies and techniques to retrieve the broach should they encounter this issue.

Authors

Institutions

Publication Details

Journal
Arthroplasty Today
Published
2025-03-15
DOI
https://doi.org/10.1016/j.artd.2025.101642
Citations
1
Primary Topic
Orthopaedic implants and arthroplasty
Type
article
Field-Weighted Citation Impact
2.01
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Fracture of the Femoral Broach Locking Mechanism and Surgical Technique to Retrieve the Broach

Friedrich Boettner, Peter P. Hsiue, Ryan Cheng, Elliot Sappey‐Marinier et al.
1 citations
Arthroplasty Today
Orthopaedic implants and arthroplasty
2.01
article

Fracture of the Femoral Broach Locking Mechanism and Surgical Technique to Retrieve the Broach

Friedrich Boettner, Peter P. Hsiue, Ryan Cheng, Elliot Sappey‐Marinier, Jeffrey A. O’Donnell, Tarik Aït Si Selmi
article en
1 citations

Abstract

The current case series reports on 4 cases of intraoperative fracture of the femoral broach locking mechanism during total hip arthroplasty and subsequent entrapment of the broach. Three of the cases were performed via the direct anterior approach and 1 was performed with the posterolateral approach. Two femoral broach locking mechanisms fractured with the use of a power broaching device and 2 occurred when using the calcar planar. The authors also discuss a minimally invasive technique to remove the broach. While intraoperative fracture of the femoral broach locking mechanism is not a common complication, surgeons should be aware of the potential causes of these complications as well as strategies and techniques to retrieve the broach should they encounter this issue.

Arthroplasty TodayVol. 32
Hospital for Special Surgery (US), Hôpital Privé Jean Mermo (FR), Centre Orthopédique Santy (FR)
Openalex Percentile: Top 18%
Orthopaedic implants and arthroplasty
2.01
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.