Sensitivity and Spatial Resolution Optimization of a High-Resolution Preclinical PET With a Unique Acquisition Method

EasyPET.3D is a preclinical positron emission tomography (PET) scanner using a unique scanning method based on two face-to-face detector modules with two axes of motion. The sensitivity and spatial resolution were optimized for mouse imaging by studying the operating parameters related to motor motion (speed and step angle), following the NEMA NU 4-2008 Standards. Moreover, the impact of the energy window and positron range on the images was assessed. The fan motor should operate at a speed of 20 full steps/s, while the fan (<inline-formula xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink"> <tex-math notation="LaTeX">${F}=0.014^{\circ }$ </tex-math></inline-formula>–0.113°) and axial (<inline-formula xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink"> <tex-math notation="LaTeX">${A}=0.9^{\circ }$ </tex-math></inline-formula>–9.0°) step angles are chosen depending on the study’s purpose. The image quality experiment demonstrated the high-resolution capability of easyPET.3D. A 200–750 keV energy window maximized the sensitivity (+200%) without significantly increasing scatter fraction (SF) (+35%). In contrast, the acquisition protocol made it difficult to conclude about the positron range effect. The feature with the most impact on the scanner’s performance is the fan motor speed. A lower fan motor speed of 20 steps/s enhanced sensitivity and spatial resolution by +122% and +60%, respectively, increased noise equivalent count rate by 155%, decreased SF by 7%, and improved recovery coefficient by +35%.

Authors

Institutions

Publication Details

Journal
IEEE Transactions on Radiation and Plasma Medical Sciences
Published
2025-03-14
DOI
https://doi.org/10.1109/trpms.2025.3546120
Citations
2
Primary Topic
Medical Imaging Techniques and Applications
Type
article
Field-Weighted Citation Impact
2.72

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Sensitivity and Spatial Resolution Optimization of a High-Resolution Preclinical PET With a Unique Acquisition Method

A.L.M. Silva, Pedro Encarnação, I.F. Castro, Ana C. Santos et al.
2 citations
IEEE Transactions on Radiation and Plasma Medical Sciences
Medical Imaging Techniques and Applications
2.72
article

Sensitivity and Spatial Resolution Optimization of a High-Resolution Preclinical PET With a Unique Acquisition Method

A.L.M. Silva, Pedro Encarnação, I.F. Castro, Ana C. Santos, J.F.C.A. Veloso, F. M. Ribeiro, P. M. M. Correia, A. Pinto
article en
2 citations

Abstract

EasyPET.3D is a preclinical positron emission tomography (PET) scanner using a unique scanning method based on two face-to-face detector modules with two axes of motion. The sensitivity and spatial resolution were optimized for mouse imaging by studying the operating parameters related to motor motion (speed and step angle), following the NEMA NU 4-2008 Standards. Moreover, the impact of the energy window and positron range on the images was assessed. The fan motor should operate at a speed of 20 full steps/s, while the fan (<inline-formula xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink"> <tex-math notation="LaTeX">${F}=0.014^{\circ }$ </tex-math></inline-formula>–0.113°) and axial (<inline-formula xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink"> <tex-math notation="LaTeX">${A}=0.9^{\circ }$ </tex-math></inline-formula>–9.0°) step angles are chosen depending on the study’s purpose. The image quality experiment demonstrated the high-resolution capability of easyPET.3D. A 200–750 keV energy window maximized the sensitivity (+200%) without significantly increasing scatter fraction (SF) (+35%). In contrast, the acquisition protocol made it difficult to conclude about the positron range effect. The feature with the most impact on the scanner’s performance is the fan motor speed. A lower fan motor speed of 20 steps/s enhanced sensitivity and spatial resolution by +122% and +60%, respectively, increased noise equivalent count rate by 155%, decreased SF by 7%, and improved recovery coefficient by +35%.

IEEE Transactions on Radiation and Plasma Medical SciencesVol. 9(7)
University of Aveiro (PT), University of Coimbra (PT)
Fundação para a Ciência e a Tecnologia
Openalex Percentile: Top 12%
Medical Imaging Techniques and Applications
2.72
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.