Evaluating the Performance of the New Sysmex XR-Series Haematology Analyser: A Comparative Study with the Sysmex XN-Series
The objective of this study was to assess the performance characteristics of the new automated haematology analyser from Sysmex Corporation, the Sysmex XR-Series, compare its performance to the Sysmex XN-Series through method comparison, and compare our results to previously published literature. Analytical performance of the new Sysmex XR-20 consisting of precision, bias, and total error, a method comparison with the Sysmex XN-2000, and the flagging performance evaluation were conducted on a Sysmex XR-20 analyser in the AZ Sint-Lucas Hospital (Bruges, Belgium) several months before its launch in Europe. We conclude that the Sysmex XR-Series is an excellent successor to the Sysmex XN-Series for routine haematology analysis. Analytical performance and flagging efficiency are comparable to the Sysmex XN-analyser.