269 Characterizing the Complication Profile of Spinal Robotic Systems: A MAUDE Analysis of Device Failures and Associated Complications by Device Company and Brand

INTRODUCTION: Pedicle screw placement in spinal surgery has evolved to incorporate increased use of technology for both preoperative planning and intraoperative guidance. In recent years, robotic guidance has been used to increase the precision, accuracy, and overall patient safety of spinal column fixation via screw placement. METHODS: The Manufacturer and User Facility Device Experience (MAUDE) was queried for adverse events involving the most widely used spinal robotic devices from inception to November 3rd, 2023. Entries were tabulated by assigning categorical variables based on event information. RESULTS: The query yielded 339 unique robotic failure events. The most frequent complications in robotic-assisted spinal surgeries were imprecision (73.6%), software errors (13.7%), and damaged or missing screws (4.9%). Most reported events occurred intraoperatively after the initial incision (78.5%) [compared to preoperatively or postoperatively]. The Mazor X™ system was the most represented spinal robotic platform in terms of complications, accounting for 71.6% of reported robotic failure events. Most reported errors were resolved intraoperatively or continued through the employment of alternative imaging methods and, overall, had minimal clinical significance. With respect to robot-assisted pedicle screw placement, medial breaches were 1.7x more common than lateral breaches across all robotic systems. Finally, regarding the impact of device failures on perioperative timing, we found that 10.3% of adverse events resulted in delays greater than 1 hour. CONCLUSIONS: Ultimately, imprecision and unresponsive software were found to be two of the most frequent complications of robot-assisted spine surgery. Though most reported errors were circumvented intraoperatively through conventional, non-robotic imaging/guidance methods, 10.3% of adverse events resulted in procedural delays of greater than 1 hour.

Authors

Publication Details

Journal
Neurosurgery
Published
2025-03-14
DOI
https://doi.org/10.1227/neu.0000000000003360_269
Primary Topic
Biomedical Ethics and Regulation
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

269 Characterizing the Complication Profile of Spinal Robotic Systems: A MAUDE Analysis of Device Failures and Associated Complications by Device Company and Brand

Neal Patel, Ronald Sahyouni, Nolan J. Brown, Zach Pennington et al.
Neurosurgery
Biomedical Ethics and Regulation
article

269 Characterizing the Complication Profile of Spinal Robotic Systems: A MAUDE Analysis of Device Failures and Associated Complications by Device Company and Brand

Neal Patel, Ronald Sahyouni, Nolan J. Brown, Zach Pennington, Julian Gendreau, Martin H. Pham
article en

Abstract

INTRODUCTION: Pedicle screw placement in spinal surgery has evolved to incorporate increased use of technology for both preoperative planning and intraoperative guidance. In recent years, robotic guidance has been used to increase the precision, accuracy, and overall patient safety of spinal column fixation via screw placement. METHODS: The Manufacturer and User Facility Device Experience (MAUDE) was queried for adverse events involving the most widely used spinal robotic devices from inception to November 3rd, 2023. Entries were tabulated by assigning categorical variables based on event information. RESULTS: The query yielded 339 unique robotic failure events. The most frequent complications in robotic-assisted spinal surgeries were imprecision (73.6%), software errors (13.7%), and damaged or missing screws (4.9%). Most reported events occurred intraoperatively after the initial incision (78.5%) [compared to preoperatively or postoperatively]. The Mazor X™ system was the most represented spinal robotic platform in terms of complications, accounting for 71.6% of reported robotic failure events. Most reported errors were resolved intraoperatively or continued through the employment of alternative imaging methods and, overall, had minimal clinical significance. With respect to robot-assisted pedicle screw placement, medial breaches were 1.7x more common than lateral breaches across all robotic systems. Finally, regarding the impact of device failures on perioperative timing, we found that 10.3% of adverse events resulted in delays greater than 1 hour. CONCLUSIONS: Ultimately, imprecision and unresponsive software were found to be two of the most frequent complications of robot-assisted spine surgery. Though most reported errors were circumvented intraoperatively through conventional, non-robotic imaging/guidance methods, 10.3% of adverse events resulted in procedural delays of greater than 1 hour.

NeurosurgeryVol. 71(Supplement_1)
Openalex Percentile: Top 94%
Biomedical Ethics and Regulation
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.