1223 Enhancing Cervical Disk Arthroplasty–Tribological Superiority of Surface Modification using Diffusion Hardened Titanium (DHT) Implants

INTRODUCTION: Commercially available cervical arthroplasty devices vary by composition and mechanical properties. Metal-on-metal (MoM) endplate articulation is associated with significantly higher odds of long-term neurological success, compared to metal-UHMWPE. Current MoM devices use either 90% Ti-6Al-4V and 10% Ti-Carbide (Ti64/TiC) composite/Cobalt-Chromium-Molybdenum alloy (CoCrMo), which are expensive. METHODS: Diffusion hardened titanium (DHT) was obtained by treatment of titanium discs at high temperature by four vendors. Four samples from each DHT were tested in tribometer using bovine calf serum. The pin-on-disc wear was run with dynamic partner of 6mm diameter, at 30N normal load, 1Hz frequency for total linear stroke length of 10mm for 1 hour. Wear rates were evaluated using stylus profilometer (20µm radius diamond-spherical tip). RESULTS: The average CoF between pin and disc of same materials were: 0.2303±0.0023 (TC4), 0.1240±0.0013(E30), 0.1850±0.0008(E80) and 0.2211±0.0204(APT). The CoF for controls were higher: 0.2465±0.0022(CoCrMo) and 0.2445±0.0006(Ti64/TiC). Wear-track depths after pin-on-disc tests were significantly lesser than TDH case depths: 2.223µm(TC4), 0.2385µm(X30), 0.248µm(X80), and 0.41µm(APT). While CoCrMo had comparable 0.619µm wear-track depth, Ti64/TiC was high at 35.75µm. The average volumetric wear rates of disc and pin combinations(in mm3/Nm) respectively were: 4.06x10-7 and 2985x10-7(TC4); 1.95x10-7 and 1.12x10-7(X30); 2.34x10-7 and 6.57x10-7(X80); 1.75x10-7 and 43.3x10-7(APT), 2.91x10-7 and 32.0x10-7(CoCrMo); 1120x10-7 and 69.9x10-7(Ti64/TiC). DHT sample discs have consistent CoF and wear rates against the pins of same materials, comparable to or better than the controls. E30, E80 and APT exhibited consistently excellent friction and wear protection between both discs and pins. CONCLUSIONS: DHT discs can outperform current controls Ti64/TiC and CoCrMo on wear and friction profiles. Therefore, DHT is promising surface modification technique to improve tribological properties of arthroplasty devices.

Authors

Publication Details

Journal
Neurosurgery
Published
2025-03-14
DOI
https://doi.org/10.1227/neu.0000000000003360_1223
Primary Topic
Spine and Intervertebral Disc Pathology
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

1223 Enhancing Cervical Disk Arthroplasty–Tribological Superiority of Surface Modification using Diffusion Hardened Titanium (DHT) Implants

Mohamad Bydon, Karim Rizwan Nathani, Asimina Dominari, Sufyan Ibrahim et al.
Neurosurgery
Spine and Intervertebral Disc Pathology
article

1223 Enhancing Cervical Disk Arthroplasty–Tribological Superiority of Surface Modification using Diffusion Hardened Titanium (DHT) Implants

Mohamad Bydon, Karim Rizwan Nathani, Asimina Dominari, Sufyan Ibrahim, Brett A. Freedman
article en

Abstract

INTRODUCTION: Commercially available cervical arthroplasty devices vary by composition and mechanical properties. Metal-on-metal (MoM) endplate articulation is associated with significantly higher odds of long-term neurological success, compared to metal-UHMWPE. Current MoM devices use either 90% Ti-6Al-4V and 10% Ti-Carbide (Ti64/TiC) composite/Cobalt-Chromium-Molybdenum alloy (CoCrMo), which are expensive. METHODS: Diffusion hardened titanium (DHT) was obtained by treatment of titanium discs at high temperature by four vendors. Four samples from each DHT were tested in tribometer using bovine calf serum. The pin-on-disc wear was run with dynamic partner of 6mm diameter, at 30N normal load, 1Hz frequency for total linear stroke length of 10mm for 1 hour. Wear rates were evaluated using stylus profilometer (20µm radius diamond-spherical tip). RESULTS: The average CoF between pin and disc of same materials were: 0.2303±0.0023 (TC4), 0.1240±0.0013(E30), 0.1850±0.0008(E80) and 0.2211±0.0204(APT). The CoF for controls were higher: 0.2465±0.0022(CoCrMo) and 0.2445±0.0006(Ti64/TiC). Wear-track depths after pin-on-disc tests were significantly lesser than TDH case depths: 2.223µm(TC4), 0.2385µm(X30), 0.248µm(X80), and 0.41µm(APT). While CoCrMo had comparable 0.619µm wear-track depth, Ti64/TiC was high at 35.75µm. The average volumetric wear rates of disc and pin combinations(in mm3/Nm) respectively were: 4.06x10-7 and 2985x10-7(TC4); 1.95x10-7 and 1.12x10-7(X30); 2.34x10-7 and 6.57x10-7(X80); 1.75x10-7 and 43.3x10-7(APT), 2.91x10-7 and 32.0x10-7(CoCrMo); 1120x10-7 and 69.9x10-7(Ti64/TiC). DHT sample discs have consistent CoF and wear rates against the pins of same materials, comparable to or better than the controls. E30, E80 and APT exhibited consistently excellent friction and wear protection between both discs and pins. CONCLUSIONS: DHT discs can outperform current controls Ti64/TiC and CoCrMo on wear and friction profiles. Therefore, DHT is promising surface modification technique to improve tribological properties of arthroplasty devices.

NeurosurgeryVol. 71(Supplement_1)
Openalex Percentile: Top 94%
Spine and Intervertebral Disc Pathology
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.