An Efficient Method for Calibrating Piezoresistive Pressure Sensor
This paper presents an efficient calibration method for piezoresistive pressure sensors. We applied this method to calibrate 35 pressure sensors integrated into an e-textile insole, utilizing machine-measured pressure-resistance pairs within the pressure range of 1–10 N as reference values for error assessment. The efficient calibration process required only 15 minutes, resulting in an average relative error of 14.4%. In contrast, traditional calibration methods necessitated more than two hours, achieving a relative error of 12.1%. This method effectively addresses the calibration challenges of piezoresistive pressure sensors within practical device and time cost constraints, holding significant implications for the real-world application of pressure sensor products that require frequent calibration.
Authors
- Zhanqing Pu (ORCID: https://orcid.org/0009-0009-0751-2037)
- Daisuke Miyata (ORCID: https://orcid.org/0000-0002-6785-2342)
- Lei Jing (ORCID: https://orcid.org/0000-0002-1181-2536)
Institutions
- University of Aizu (JP)
Publication Details
- Published
- 2024-11-19
- DOI
- https://doi.org/10.23919/e-textiles63767.2024.10914233
- Citations
- 2
- Primary Topic
- Sensor Technology and Measurement Systems
- Type
- article
- Field-Weighted Citation Impact
- 0.70