<ns3:p>The McLaren Proton Therapy Center functions as a multi-room, clinically active cancer treatment center. Beam position, size, and current measurements along the transfer lines have been utilized together with simulations to validate computer models for the Center’s transfer lines and associated beam parameters. Beam profile monitor (BPM) data collected along the beamlines provided detailed beam measurements. A computer tool, utilizing nonlinear constrained optimization programs that are integrated with a beam envelope code, was used to find detailed beam phase space parameters (centroids, sigma matrices) that reproduced the measured data. Quadrupole scan and steerer scan experiments have been used to validate the calibration and alignment of these devices, as well as to independently confirm beam parameters at selected locations. Beam current measurements along the beamlines were used to infer transmission efficiencies and then compared to multi-particle simulations in order to identify beam loss locations. The work resulted in an improved understanding of the beams and beam transfer lines to support further beamline optimization and design activities. The experimental and simulation methodologies, tools, and results and their interpretations should be applicable to many types of operating accelerator facilities.</ns3:p>

Authors

Publication Details

Journal
Nuclear Science and Technology Open Research
Published
2025-03-13
DOI
https://doi.org/10.12688/nuclscitechnolopenres.17622.2
Primary Topic
Radiation Therapy and Dosimetry
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Detailed Transfer Line Beam Characteristics at an Operational Proton Therapy Center

J. Konzer, George H. Gillespie
Nuclear Science and Technology Open Research
Radiation Therapy and Dosimetry
article

Detailed Transfer Line Beam Characteristics at an Operational Proton Therapy Center

J. Konzer, George H. Gillespie
article en

Abstract

<ns3:p>The McLaren Proton Therapy Center functions as a multi-room, clinically active cancer treatment center. Beam position, size, and current measurements along the transfer lines have been utilized together with simulations to validate computer models for the Center’s transfer lines and associated beam parameters. Beam profile monitor (BPM) data collected along the beamlines provided detailed beam measurements. A computer tool, utilizing nonlinear constrained optimization programs that are integrated with a beam envelope code, was used to find detailed beam phase space parameters (centroids, sigma matrices) that reproduced the measured data. Quadrupole scan and steerer scan experiments have been used to validate the calibration and alignment of these devices, as well as to independently confirm beam parameters at selected locations. Beam current measurements along the beamlines were used to infer transmission efficiencies and then compared to multi-particle simulations in order to identify beam loss locations. The work resulted in an improved understanding of the beams and beam transfer lines to support further beamline optimization and design activities. The experimental and simulation methodologies, tools, and results and their interpretations should be applicable to many types of operating accelerator facilities.</ns3:p>

Nuclear Science and Technology Open ResearchVol. 2
Openalex Percentile: Top 93%
Radiation Therapy and Dosimetry
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.