Accurate diagnosis of bone degradation-induced prosthesis loosening using harmonic vibration analysis: An experimental study with a simplified model

Abstract. The Osseointegration Prosthetic Limb is a transfemoral implant that attaches the prosthesis directly to the bone, providing greater comfort than a traditional socket. However, transferring load to an implant can induce stress shielding and bone resorption, potentially leading to loosening and pain. This experimental study investigates the nonlinear vibration response of a simplified femur-implant model as the basis for a non-invasive strategy to monitor implant loosening. Three femur-implant assemblies were constructed with prescribed levels of interference fit between the bone and implant. To mimic bone degradation due to stress shielding, bone thickness was reduced by material removal, which led to a reduction of contact stability. In the vibration analysis, a static longitudinal gait force was applied at the distal end of the implant to preload it with a load representative of the compression experienced during walking. Subsequently, the model was excited at the resonant frequency of the first torsional vibration mode to assess interfacial contact. Slippage at the interface occurs when the interfacial shear stress exceeds a prescribed critical threshold, leading to the generation of higher-order harmonics in the vibration response. The evolution of these harmonics was analysed in relation to the magnitude of torsional excitation and contact stability. Results suggest that the presence of the higher-order harmonic serves as a reliable indicator of interfacial slippage, and the corresponding load level provides a quantitative index of the implant's load-bearing capacity. These findings demonstrate the feasibility of using nonlinear vibration analysis, as a non-invasive means of assessing the criticality of the bone-implant interface.

Authors

Institutions

Publication Details

Journal
Materials research proceedings
Published
2025-03-13
DOI
https://doi.org/10.21741/9781644903513-31
Primary Topic
Orthopaedic implants and arthroplasty
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Accurate diagnosis of bone degradation-induced prosthesis loosening using harmonic vibration analysis: An experimental study with a simplified model

Qi Zhou
Materials research proceedings
Orthopaedic implants and arthroplasty
article

Accurate diagnosis of bone degradation-induced prosthesis loosening using harmonic vibration analysis: An experimental study with a simplified model

Qi Zhou
article en

Abstract

Abstract. The Osseointegration Prosthetic Limb is a transfemoral implant that attaches the prosthesis directly to the bone, providing greater comfort than a traditional socket. However, transferring load to an implant can induce stress shielding and bone resorption, potentially leading to loosening and pain. This experimental study investigates the nonlinear vibration response of a simplified femur-implant model as the basis for a non-invasive strategy to monitor implant loosening. Three femur-implant assemblies were constructed with prescribed levels of interference fit between the bone and implant. To mimic bone degradation due to stress shielding, bone thickness was reduced by material removal, which led to a reduction of contact stability. In the vibration analysis, a static longitudinal gait force was applied at the distal end of the implant to preload it with a load representative of the compression experienced during walking. Subsequently, the model was excited at the resonant frequency of the first torsional vibration mode to assess interfacial contact. Slippage at the interface occurs when the interfacial shear stress exceeds a prescribed critical threshold, leading to the generation of higher-order harmonics in the vibration response. The evolution of these harmonics was analysed in relation to the magnitude of torsional excitation and contact stability. Results suggest that the presence of the higher-order harmonic serves as a reliable indicator of interfacial slippage, and the corresponding load level provides a quantitative index of the implant's load-bearing capacity. These findings demonstrate the feasibility of using nonlinear vibration analysis, as a non-invasive means of assessing the criticality of the bone-implant interface.

Materials research proceedingsVol. 50
Monash University (AU)
Openalex Percentile: Top 92%
Orthopaedic implants and arthroplasty
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Accurate diagnosis of bone degradation-induced prosthesis loosening using harmonic vibration analysis: An experimental study with a simplified model — Qi Zhou · Materials research proceedings (2025) | TGRS Research Map | TGRS