A Posterior Segment Ocular Microsurgical Robot With a Decoupling RCM Mechanism Based on a Minimized Internal Constraint Force Optimization Method

This paper presents a leader-follower robotic system featuring a novel 4-degree-of-freedom (4-DOF) Remote Center of Motion (RCM) mechanism, tailored to address the limitations associated with traditional posterior segment ocular microsurgery. The proposed 4-DOF mechanism employs parallelogram motion replication to relocate the bulky instrument insertion drive from the end-effector to the proximal linkage on the base, minimizing obstruction to the microscope’s field of view and the surgical environment. The mechanism’s orthogonal or coincident arranged degrees of freedom, paired with independent drive units, facilitate calibration and enhance control accuracy. A minimized internal constraint force optimization method was proposed to improve RCM point stability and tip positioning accuracy. The mechanism’s configuration and parameters were optimized through an analytical mechanics model, effectively reducing the constraint force on the internal components under the same external loads, thereby minimizing deformation and maintaining accuracy. To enhance robot compatibility, modular surgical instruments with a quick-change coupler were developed based on the analysis of traditional instruments’ characteristics. The prototype was built and kinematically calibrated to reduce manufacturing and assembly errors and thus improve accuracy. Following kinematic calibration to minimize manufacturing and assembly errors, experimental validation revealed positioning accuracies of <inline-formula xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink"> <tex-math notation="LaTeX">$49~\pm ~23~\mu $ </tex-math></inline-formula>m and <inline-formula xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink"> <tex-math notation="LaTeX">$22~\pm ~13~\mu $ </tex-math></inline-formula>m, repeatabilities of <inline-formula xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink"> <tex-math notation="LaTeX">$25~\pm ~10~\mu $ </tex-math></inline-formula>m and <inline-formula xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink"> <tex-math notation="LaTeX">$9~\pm ~4~\mu $ </tex-math></inline-formula>m, and RCM deviations of <inline-formula xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink"> <tex-math notation="LaTeX">$13~\pm ~10~\mu $ </tex-math></inline-formula>m and <inline-formula xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink"> <tex-math notation="LaTeX">$18~\pm ~11~\mu $ </tex-math></inline-formula>m on the X-Z and Y-Z planes, respectively. The cannulation experiment further demonstrates the prototype’s potential for robot-assisted vitreoretinal microsurgery.

Authors

Institutions

Publication Details

Journal
IEEE Transactions on Medical Robotics and Bionics
Published
2025-03-12
DOI
https://doi.org/10.1109/tmrb.2025.3550643
Citations
2
Primary Topic
Soft Robotics and Applications
Type
article
Field-Weighted Citation Impact
1.35

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

A Posterior Segment Ocular Microsurgical Robot With a Decoupling RCM Mechanism Based on a Minimized Internal Constraint Force Optimization Method

Chaoyang Shi, Feiyang Chen, Dunfa Long, Zhi Li et al.
2 citations
IEEE Transactions on Medical Robotics and Bionics
Soft Robotics and Applications
1.35
article

A Posterior Segment Ocular Microsurgical Robot With a Decoupling RCM Mechanism Based on a Minimized Internal Constraint Force Optimization Method

Chaoyang Shi, Feiyang Chen, Dunfa Long, Zhi Li, Kaifeng Wang
article en
2 citations

Abstract

This paper presents a leader-follower robotic system featuring a novel 4-degree-of-freedom (4-DOF) Remote Center of Motion (RCM) mechanism, tailored to address the limitations associated with traditional posterior segment ocular microsurgery. The proposed 4-DOF mechanism employs parallelogram motion replication to relocate the bulky instrument insertion drive from the end-effector to the proximal linkage on the base, minimizing obstruction to the microscope’s field of view and the surgical environment. The mechanism’s orthogonal or coincident arranged degrees of freedom, paired with independent drive units, facilitate calibration and enhance control accuracy. A minimized internal constraint force optimization method was proposed to improve RCM point stability and tip positioning accuracy. The mechanism’s configuration and parameters were optimized through an analytical mechanics model, effectively reducing the constraint force on the internal components under the same external loads, thereby minimizing deformation and maintaining accuracy. To enhance robot compatibility, modular surgical instruments with a quick-change coupler were developed based on the analysis of traditional instruments’ characteristics. The prototype was built and kinematically calibrated to reduce manufacturing and assembly errors and thus improve accuracy. Following kinematic calibration to minimize manufacturing and assembly errors, experimental validation revealed positioning accuracies of <inline-formula xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink"> <tex-math notation="LaTeX">$49~\pm ~23~\mu $ </tex-math></inline-formula>m and <inline-formula xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink"> <tex-math notation="LaTeX">$22~\pm ~13~\mu $ </tex-math></inline-formula>m, repeatabilities of <inline-formula xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink"> <tex-math notation="LaTeX">$25~\pm ~10~\mu $ </tex-math></inline-formula>m and <inline-formula xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink"> <tex-math notation="LaTeX">$9~\pm ~4~\mu $ </tex-math></inline-formula>m, and RCM deviations of <inline-formula xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink"> <tex-math notation="LaTeX">$13~\pm ~10~\mu $ </tex-math></inline-formula>m and <inline-formula xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink"> <tex-math notation="LaTeX">$18~\pm ~11~\mu $ </tex-math></inline-formula>m on the X-Z and Y-Z planes, respectively. The cannulation experiment further demonstrates the prototype’s potential for robot-assisted vitreoretinal microsurgery.

IEEE Transactions on Medical Robotics and BionicsVol. 7(2)
Tianjin University (CN)
National Natural Science Foundation of China
Climate action
Openalex Percentile: Top 22%
Soft Robotics and Applications
1.35
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.