Performance evaluation of Rayto RT-7600Vet hematology analyzer in side-by-side comparison with manual hematological methods for apparently healthy Cholistani cattle blood

The present study is the first from Pakistan being reported with an objective to assess performance of Rayto RT-7600Vet hematology analyzer (HA) for Cholistani cattle blood (n = 134), in comparison to the manual hematological methods. The four hematological attributes viz. total erythrocyte count (TEC), hemoglobin (Hb), packed cell volume (PCV) and platelet count (PLT) were deduced through HA (A) and manual (M) methods. Various statistical tests were implied to ascertain level of interrelationship, accuracy and level of agreement between the two methods. All attributes attained through manual methods had high positive, leptokurtic distribution (having many outliers) except for PLT-M and PCV-M. The coefficient of variation for attributes attained through HA and manual methods ranged from 16-24% and from 16-59%, respectively. Comparison between the overall results revealed that all the studied attributes, except TEC, were significantly (P≤0.05) different for both methods. A weak relationship was noticed between the attributes attained through two methods as indicated by weak r-values and adjusted r-square values. The reliability level of estimating Hb and PCV had highest intraclass correlation coefficient value of 0.722 and 0.555 for average measures, respectively. However, accuracy level, as determined through Lin's concordance correlation coefficient was highest for TEC (0.9504) for both analytical methods. Poor level of agreement, in general, was shown for the two methods of analysis regarding all four hematological attributes through Bland and Altman test. In conclusion, the Rayto RT-7600Vet) may present data having higher skewness, kurtosis, and CV%, however, they are valid for multi-species hematological analysis. Caution must however, be taken in interpreting their results with corrected reference intervals and CV% for each machine and for each tested attribute.

Authors

Institutions

Publication Details

Journal
PLoS ONE
Published
2025-03-11
DOI
https://doi.org/10.1371/journal.pone.0302617
Citations
1
Primary Topic
Clinical Laboratory Practices and Quality Control
Type
article
Field-Weighted Citation Impact
1.16
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Performance evaluation of Rayto RT-7600Vet hematology analyzer in side-by-side comparison with manual hematological methods for apparently healthy Cholistani cattle blood

Iram Fatima, Umer Farooq, Zia ur Rehman, Musadiq İdris et al.
1 citations
PLoS ONE
Clinical Laboratory Practices and Quality Control
1.16
article

Performance evaluation of Rayto RT-7600Vet hematology analyzer in side-by-side comparison with manual hematological methods for apparently healthy Cholistani cattle blood

Iram Fatima, Umer Farooq, Zia ur Rehman, Musadiq İdris, Shagufta Nasreen, Haroon Rashid, Mushtaq Hussain Lashari, Farah Laraib, Rubaisha Ameer, Maryam Chauhdary
article en
1 citations

Abstract

The present study is the first from Pakistan being reported with an objective to assess performance of Rayto RT-7600Vet hematology analyzer (HA) for Cholistani cattle blood (n = 134), in comparison to the manual hematological methods. The four hematological attributes viz. total erythrocyte count (TEC), hemoglobin (Hb), packed cell volume (PCV) and platelet count (PLT) were deduced through HA (A) and manual (M) methods. Various statistical tests were implied to ascertain level of interrelationship, accuracy and level of agreement between the two methods. All attributes attained through manual methods had high positive, leptokurtic distribution (having many outliers) except for PLT-M and PCV-M. The coefficient of variation for attributes attained through HA and manual methods ranged from 16-24% and from 16-59%, respectively. Comparison between the overall results revealed that all the studied attributes, except TEC, were significantly (P≤0.05) different for both methods. A weak relationship was noticed between the attributes attained through two methods as indicated by weak r-values and adjusted r-square values. The reliability level of estimating Hb and PCV had highest intraclass correlation coefficient value of 0.722 and 0.555 for average measures, respectively. However, accuracy level, as determined through Lin's concordance correlation coefficient was highest for TEC (0.9504) for both analytical methods. Poor level of agreement, in general, was shown for the two methods of analysis regarding all four hematological attributes through Bland and Altman test. In conclusion, the Rayto RT-7600Vet) may present data having higher skewness, kurtosis, and CV%, however, they are valid for multi-species hematological analysis. Caution must however, be taken in interpreting their results with corrected reference intervals and CV% for each machine and for each tested attribute.

PLoS ONEVol. 20(3)
Islamia University of Bahawalpur (PK)
No poverty
Openalex Percentile: Top 25%
Clinical Laboratory Practices and Quality Control
1.16
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.