Phenolic Preservatives Are Not the Sole Cause of Eosinophilic Infiltration at Infusion Pump Sites

<b><i>Background:</i></b> Skin reactions and discomfort associated with insulin infusion pumps limit user adherence. A recent histopathological study by Kalus et al. (DERMIS study) reported increased eosinophilic infiltration and imputed an inflammatory response to an allergen delivered at the catheter tip. This finding might explain the pruritus reported by pump users. As eosinophils migrate to inflammatory foci, primarily due to IL-5 and CCL11, we aimed to evaluate insulin phenolic preservative (IPP) as a potential allergen in vitro and assess tissue eosinophilic infiltration in vivo. <b><i>Methods:</i></b> Histopathological evaluations for eosinophil recruitment were performed over 1 week following IPP infusions in swine tissue. Additional histopathological investigations of eosinophilic infiltration were conducted using three commercial glucose sensors implanted in swine for up to 3 weeks. <b><i>Results:</i></b> Eosinophilic infiltration in the dermis and subcutaneous tissue was observed following saline and IPP infusion and at glucose sensor implantation at all time points examined. In vitro studies revealed IPP eosinophil cytotoxicity. However, neither CCL11 nor IL-5 was detected in any of the tested tissue cells after IPP treatment. <b><i>Conclusion:</i></b> These findings suggest that IPP is not the only triggering allergen, as IPP did not induce eosinophils in vitro, while glucose sensors also indicated increased eosinophilic infiltration. Therefore, factors other than IPP trigger eosinophil recruitment to insulin infusion pump sets.

Authors

Institutions

Publication Details

Journal
Diabetes Technology & Therapeutics
Published
2025-03-11
DOI
https://doi.org/10.1089/dia.2025.0043
Primary Topic
Eosinophilic Disorders and Syndromes
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Phenolic Preservatives Are Not the Sole Cause of Eosinophilic Infiltration at Infusion Pump Sites

Ulrike Klueh, Kenneth Wood, Li Mao, Priscila Silva Cunegundes
Diabetes Technology & Therapeutics
Eosinophilic Disorders and Syndromes
article

Phenolic Preservatives Are Not the Sole Cause of Eosinophilic Infiltration at Infusion Pump Sites

Ulrike Klueh, Kenneth Wood, Li Mao, Priscila Silva Cunegundes
article en

Abstract

<b><i>Background:</i></b> Skin reactions and discomfort associated with insulin infusion pumps limit user adherence. A recent histopathological study by Kalus et al. (DERMIS study) reported increased eosinophilic infiltration and imputed an inflammatory response to an allergen delivered at the catheter tip. This finding might explain the pruritus reported by pump users. As eosinophils migrate to inflammatory foci, primarily due to IL-5 and CCL11, we aimed to evaluate insulin phenolic preservative (IPP) as a potential allergen in vitro and assess tissue eosinophilic infiltration in vivo. <b><i>Methods:</i></b> Histopathological evaluations for eosinophil recruitment were performed over 1 week following IPP infusions in swine tissue. Additional histopathological investigations of eosinophilic infiltration were conducted using three commercial glucose sensors implanted in swine for up to 3 weeks. <b><i>Results:</i></b> Eosinophilic infiltration in the dermis and subcutaneous tissue was observed following saline and IPP infusion and at glucose sensor implantation at all time points examined. In vitro studies revealed IPP eosinophil cytotoxicity. However, neither CCL11 nor IL-5 was detected in any of the tested tissue cells after IPP treatment. <b><i>Conclusion:</i></b> These findings suggest that IPP is not the only triggering allergen, as IPP did not induce eosinophils in vitro, while glucose sensors also indicated increased eosinophilic infiltration. Therefore, factors other than IPP trigger eosinophil recruitment to insulin infusion pump sets.

Diabetes Technology & TherapeuticsVol. 27(8)
Wayne State University (US)
Zero hunger
Openalex Percentile: Top 93%
Eosinophilic Disorders and Syndromes
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Phenolic Preservatives Are Not the Sole Cause of Eosinophilic Infiltration at Infusion Pump Sites — Ulrike Klueh, Kenneth Wood, et al. · Diabetes Technology & Therapeutics (2025) | TGRS Research Map | TGRS