Highly-Sensitive Polycrystalline Diamond X-Ray Dosimeter With Hydrogen Termination Contacts

This letter reports the polycrystalline diamond (PCD) dosimeter with near ideal linearity and high sensitivity to meet the application requirements of medical X-ray dose measurement for low cost, large size, and high performance. The device is fabricated by using hydrogen termination modulation ohmic contact and high-quality materials with preferred (111) and (311) crystal orientation, large grain size and low impurity content. The linearity, specific sensitivity, and current gain obtained from experimental measurements are 0.999, <inline-formula xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink"> <tex-math notation="LaTeX">$1.482~\mu $ </tex-math></inline-formula>C<inline-formula xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink"> <tex-math notation="LaTeX">$\cdot $ </tex-math></inline-formula>Gy<inline-formula xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink"> <tex-math notation="LaTeX">$^{-{1}}\cdot $ </tex-math></inline-formula>mm<inline-formula xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink"> <tex-math notation="LaTeX">$^{-{3}}$ </tex-math></inline-formula> and 5.76 at 300 V (1 V/<inline-formula xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink"> <tex-math notation="LaTeX">$\mu $ </tex-math></inline-formula>m), respectively. Furthermore, the signal-to-noise ratio that meets the requirements of International Atomic Energy Agency is greater than <inline-formula xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink"> <tex-math notation="LaTeX">$10^{{4}}$ </tex-math></inline-formula> at 5 to 300 V, and the current fluctuations under dynamic voltage switch test condition of more than 30 cycle (2 s on-state and 1 s off-state) is only 0.241% at 200 V under dose rate of 6.184 Gy/min. This study indicates that PCD dosimeter fabricated by optimizing contact structure and material properties has a great potential for medical dose measurement.

Authors

Institutions

Publication Details

Journal
IEEE Electron Device Letters
Published
2025-03-11
DOI
https://doi.org/10.1109/led.2025.3549838
Citations
1
Primary Topic
Diamond and Carbon-based Materials Research
Type
article
Field-Weighted Citation Impact
0.52

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Highly-Sensitive Polycrystalline Diamond X-Ray Dosimeter With Hydrogen Termination Contacts

Xiaoping Ouyang, Zeyang Ren, Weimin Bao, Chang Liu et al.
1 citations
IEEE Electron Device Letters
Diamond and Carbon-based Materials Research
0.52
article

Highly-Sensitive Polycrystalline Diamond X-Ray Dosimeter With Hydrogen Termination Contacts

Xiaoping Ouyang, Zeyang Ren, Weimin Bao, Chang Liu, Kai Su, Jinfeng Zhang, Jincheng Zhang, Yachao Zhang, Yue Hao, Wei Wang
article en
1 citations

Abstract

This letter reports the polycrystalline diamond (PCD) dosimeter with near ideal linearity and high sensitivity to meet the application requirements of medical X-ray dose measurement for low cost, large size, and high performance. The device is fabricated by using hydrogen termination modulation ohmic contact and high-quality materials with preferred (111) and (311) crystal orientation, large grain size and low impurity content. The linearity, specific sensitivity, and current gain obtained from experimental measurements are 0.999, <inline-formula xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink"> <tex-math notation="LaTeX">$1.482~\mu $ </tex-math></inline-formula>C<inline-formula xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink"> <tex-math notation="LaTeX">$\cdot $ </tex-math></inline-formula>Gy<inline-formula xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink"> <tex-math notation="LaTeX">$^{-{1}}\cdot $ </tex-math></inline-formula>mm<inline-formula xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink"> <tex-math notation="LaTeX">$^{-{3}}$ </tex-math></inline-formula> and 5.76 at 300 V (1 V/<inline-formula xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink"> <tex-math notation="LaTeX">$\mu $ </tex-math></inline-formula>m), respectively. Furthermore, the signal-to-noise ratio that meets the requirements of International Atomic Energy Agency is greater than <inline-formula xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink"> <tex-math notation="LaTeX">$10^{{4}}$ </tex-math></inline-formula> at 5 to 300 V, and the current fluctuations under dynamic voltage switch test condition of more than 30 cycle (2 s on-state and 1 s off-state) is only 0.241% at 200 V under dose rate of 6.184 Gy/min. This study indicates that PCD dosimeter fabricated by optimizing contact structure and material properties has a great potential for medical dose measurement.

IEEE Electron Device LettersVol. 46(5)
Xidian University (CN), Northwest Institute of Nuclear Technology (CN)
National Natural Science Foundation of China, China Postdoctoral Science Foundation
Affordable and clean energy
Openalex Percentile: Top 41%
Diamond and Carbon-based Materials Research
0.52
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.