Highly-Sensitive Polycrystalline Diamond X-Ray Dosimeter With Hydrogen Termination Contacts
This letter reports the polycrystalline diamond (PCD) dosimeter with near ideal linearity and high sensitivity to meet the application requirements of medical X-ray dose measurement for low cost, large size, and high performance. The device is fabricated by using hydrogen termination modulation ohmic contact and high-quality materials with preferred (111) and (311) crystal orientation, large grain size and low impurity content. The linearity, specific sensitivity, and current gain obtained from experimental measurements are 0.999, <inline-formula xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink"> <tex-math notation="LaTeX">$1.482~\mu $ </tex-math></inline-formula>C<inline-formula xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink"> <tex-math notation="LaTeX">$\cdot $ </tex-math></inline-formula>Gy<inline-formula xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink"> <tex-math notation="LaTeX">$^{-{1}}\cdot $ </tex-math></inline-formula>mm<inline-formula xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink"> <tex-math notation="LaTeX">$^{-{3}}$ </tex-math></inline-formula> and 5.76 at 300 V (1 V/<inline-formula xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink"> <tex-math notation="LaTeX">$\mu $ </tex-math></inline-formula>m), respectively. Furthermore, the signal-to-noise ratio that meets the requirements of International Atomic Energy Agency is greater than <inline-formula xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink"> <tex-math notation="LaTeX">$10^{{4}}$ </tex-math></inline-formula> at 5 to 300 V, and the current fluctuations under dynamic voltage switch test condition of more than 30 cycle (2 s on-state and 1 s off-state) is only 0.241% at 200 V under dose rate of 6.184 Gy/min. This study indicates that PCD dosimeter fabricated by optimizing contact structure and material properties has a great potential for medical dose measurement.
Authors
- Xiaoping Ouyang (ORCID: https://orcid.org/0000-0001-8474-2869)
- Zeyang Ren (ORCID: https://orcid.org/0000-0002-5503-7228)
- Weimin Bao (ORCID: https://orcid.org/0000-0002-1851-1330)
- Chang Liu (ORCID: https://orcid.org/0000-0003-1829-0838)
- Kai Su (ORCID: https://orcid.org/0000-0003-1276-5409)
- Jinfeng Zhang (ORCID: https://orcid.org/0000-0003-1771-5187)
- Jincheng Zhang (ORCID: https://orcid.org/0000-0001-7332-6704)
- Yachao Zhang (ORCID: https://orcid.org/0000-0003-1864-6953)
- Yue Hao (ORCID: https://orcid.org/0000-0002-8081-2919)
- Wei Wang (ORCID: https://orcid.org/0000-0003-1568-222X)
Institutions
- Xidian University (CN)
- Northwest Institute of Nuclear Technology (CN)
Publication Details
- Journal
- IEEE Electron Device Letters
- Published
- 2025-03-11
- DOI
- https://doi.org/10.1109/led.2025.3549838
- Citations
- 1
- Primary Topic
- Diamond and Carbon-based Materials Research
- Type
- article
- Field-Weighted Citation Impact
- 0.52
Funders
- National Natural Science Foundation of China
- China Postdoctoral Science Foundation