Implantation and Follow‐Up of an Extravascular ICD After a Challenging Transvenous System Extraction: A Case Report

The Extravascular ICD (EV-ICD) is implanted outside of the patient's heart and vascular system. The defibrillation lead is tunneled in the substernal space, very close to the surface of the heart, thus allowing for additional pacing capabilities such as Post Shock Pacing, Anti-Tachycardia Pacing (ATP), and Pause Prevention Pacing. Early studies demonstrated the safety and efficacy profile of the EV-ICD, but patients with previous or existing cardiac devices or leads were excluded. In this case report we describe the implantation of an EV-ICD in a 68-year-old male patient, who had undergone a challenging extraction of a previous transvenous system due to infection and screened negative for a subcutaneous ICD (S-ICD). Given the lack of upper and lower venous accesses, it was also necessary to implant an epicardial bipolar lead connected to a single-chamber pacemaker as a backup during extraction, which also remained in place chronically. Follow-up was uneventful through the following 10 months, with stable system parameters and no infection relapse.

Authors

Institutions

Publication Details

Journal
Pacing and Clinical Electrophysiology
Published
2025-03-08
DOI
https://doi.org/10.1111/pace.15170
Citations
1
Primary Topic
Cardiac pacing and defibrillation studies
Type
article
Field-Weighted Citation Impact
1.39
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Implantation and Follow‐Up of an Extravascular ICD After a Challenging Transvenous System Extraction: A Case Report

Gianluca Epicoco, Tomaso Bottio, Rossella Troccoli, Carlo D’Agostino et al.
1 citations
Pacing and Clinical Electrophysiology
Cardiac pacing and defibrillation studies
1.39
article

Implantation and Follow‐Up of an Extravascular ICD After a Challenging Transvenous System Extraction: A Case Report

Gianluca Epicoco, Tomaso Bottio, Rossella Troccoli, Carlo D’Agostino, Ida Pastore, Nicola Di Bari, D Carretta
article en
1 citations

Abstract

The Extravascular ICD (EV-ICD) is implanted outside of the patient's heart and vascular system. The defibrillation lead is tunneled in the substernal space, very close to the surface of the heart, thus allowing for additional pacing capabilities such as Post Shock Pacing, Anti-Tachycardia Pacing (ATP), and Pause Prevention Pacing. Early studies demonstrated the safety and efficacy profile of the EV-ICD, but patients with previous or existing cardiac devices or leads were excluded. In this case report we describe the implantation of an EV-ICD in a 68-year-old male patient, who had undergone a challenging extraction of a previous transvenous system due to infection and screened negative for a subcutaneous ICD (S-ICD). Given the lack of upper and lower venous accesses, it was also necessary to implant an epicardial bipolar lead connected to a single-chamber pacemaker as a backup during extraction, which also remained in place chronically. Follow-up was uneventful through the following 10 months, with stable system parameters and no infection relapse.

Pacing and Clinical ElectrophysiologyVol. 48(4)
Azienda Universitaria Ospedaliera Consorziale - Policlinico Bari (IT)
Good health and well-being
Openalex Percentile: Top 22%
Cardiac pacing and defibrillation studies
1.39
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.