A dosimetric comparison of helical tomotherapy treatment delivery with real-time adaption and no motion correction

This study assesses the ability of a helical tomotherapy system equipped with kV imaging and optical surface guidance to adapt to motion traces in real-time. To assess the delivery accuracy with motion, a unified testing framework was used. The average 2 %/2 mm γ-fail rates across all lung traces were 0.1 % for motion adapted and 17.4 % for no motion correction. Average 2 %/2 mm γ-fail rates across all prostate traces were 0.4 % for motion adapted and 12.2 % for no motion correction. Real-time motion adaption was shown to improve the accuracy of dose delivered to a moving phantom compared with no motion adaption. <b>MeSH Keywords:</b> Radiotherapy, image-guided; Radiation therapy, targeted.

Authors

Institutions

Publication Details

Journal
Physics and Imaging in Radiation Oncology
Published
2025-03-05
DOI
https://doi.org/10.1016/j.phro.2025.100741
Citations
8
Primary Topic
Advanced Radiotherapy Techniques
Type
article
Field-Weighted Citation Impact
15.06
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

A dosimetric comparison of helical tomotherapy treatment delivery with real-time adaption and no motion correction

Paul Keall, Jeremy Booth, C. Sengupta, Jonathan Hindmarsh et al.
8 citations
Physics and Imaging in Radiation Oncology
Advanced Radiotherapy Techniques
15.06
article

A dosimetric comparison of helical tomotherapy treatment delivery with real-time adaption and no motion correction

Paul Keall, Jeremy Booth, C. Sengupta, Jonathan Hindmarsh, Scott Crowe, Sonja Dieterich, Julia V. Johnson, J. Walsh
article en
8 citations

Abstract

This study assesses the ability of a helical tomotherapy system equipped with kV imaging and optical surface guidance to adapt to motion traces in real-time. To assess the delivery accuracy with motion, a unified testing framework was used. The average 2 %/2 mm γ-fail rates across all lung traces were 0.1 % for motion adapted and 17.4 % for no motion correction. Average 2 %/2 mm γ-fail rates across all prostate traces were 0.4 % for motion adapted and 12.2 % for no motion correction. Real-time motion adaption was shown to improve the accuracy of dose delivered to a moving phantom compared with no motion adaption. <b>MeSH Keywords:</b> Radiotherapy, image-guided; Radiation therapy, targeted.

Physics and Imaging in Radiation OncologyVol. 34
University of Technology Sydney (AU), The University of Sydney (AU), Royal Brisbane and Women's Hospital (AU), Royal North Shore Hospital (AU), University of California Davis Medical Center (US)
Openalex Percentile: Top 1%
Advanced Radiotherapy Techniques
15.06
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.