The Product Quality Research Institute elemental impurity interlaboratory study: Results and implications for industry

• We organized an interlaboratory study of elemental impurity analysis in drug products • Objectives were comparing lab equipment, preparation method, and summation approach • Simulated drug tablets and raw materials were analyzed for metals by ICP-MS, XRF • Though some elements were challenging (Hg, V), results were generally consistent • Results demonstrate that some challenges must be addressed, e.g. standard practices Pharmaceutical laboratories experienced a paradigm shift in drug product elemental impurity (EI) expectations in International Council on Harmonization of Technical Requirements for Pharmaceuticals for Human Use (ICH) Guideline Q3D and United States Pharmacopeia (USP) General Chapters <232>/<233>. These guidelines describe a risk-based approach to EI analysis. Few systematic evaluations of interlaboratory performance on EI analysis in pharmaceutics have been conducted following these guidelines. Our goal is to address key technical challenges faced by laboratories during the implementation of these regulations. We organized an interlaboratory study using standardized samples and methodology to assess sample preparation and analysis variability. Participants performed microwave-assisted acid preparation of simulated pharmaceutical products and analyzed Class 1 and 2A EI's by inductively-coupled plasma-mass spectrometry (ICP-MS). Several laboratories performed X-ray Fluorescence spectroscopy (XRF) for comparison. ICP-MS reproducibility was high both within and between laboratories, except for Hg and V. Exhaustive extraction and total digestion were generally comparable, between 87-111% for As, Cd, Co, and Pb. Total digestion exhibited lower variability than exhaustive extraction. Two types of microwave systems produced comparable results for most elements except Hg and Pb. The summation approach was comparable to direct analysis of tablets except for Hg and Cd, but summation demonstrated greater variability. XRF showed good agreement with ICP-MS and low replicate variability within labs. While the results were generally favorable, they demonstrate that some technical challenges remain to be addressed related to standardizing laboratory practices including interference correction strategies and selection of preparation methods. We discuss implications for method transfer between laboratories.

Authors

Institutions

Publication Details

Journal
Journal of Trace Elements and Minerals
Published
2025-03-09
DOI
https://doi.org/10.1016/j.jtemin.2025.100227
Primary Topic
Extraction and Separation Processes
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

The Product Quality Research Institute elemental impurity interlaboratory study: Results and implications for industry

James M. Harrington, Denise McClenathan, Donna S. Seibert, Stephen W. Erickson et al.
Journal of Trace Elements and Minerals
Extraction and Separation Processes
article

The Product Quality Research Institute elemental impurity interlaboratory study: Results and implications for industry

James M. Harrington, Denise McClenathan, Donna S. Seibert, Stephen W. Erickson, Thanh Nguyen, Glenn N. Williams
article en

Abstract

• We organized an interlaboratory study of elemental impurity analysis in drug products • Objectives were comparing lab equipment, preparation method, and summation approach • Simulated drug tablets and raw materials were analyzed for metals by ICP-MS, XRF • Though some elements were challenging (Hg, V), results were generally consistent • Results demonstrate that some challenges must be addressed, e.g. standard practices Pharmaceutical laboratories experienced a paradigm shift in drug product elemental impurity (EI) expectations in International Council on Harmonization of Technical Requirements for Pharmaceuticals for Human Use (ICH) Guideline Q3D and United States Pharmacopeia (USP) General Chapters <232>/<233>. These guidelines describe a risk-based approach to EI analysis. Few systematic evaluations of interlaboratory performance on EI analysis in pharmaceutics have been conducted following these guidelines. Our goal is to address key technical challenges faced by laboratories during the implementation of these regulations. We organized an interlaboratory study using standardized samples and methodology to assess sample preparation and analysis variability. Participants performed microwave-assisted acid preparation of simulated pharmaceutical products and analyzed Class 1 and 2A EI's by inductively-coupled plasma-mass spectrometry (ICP-MS). Several laboratories performed X-ray Fluorescence spectroscopy (XRF) for comparison. ICP-MS reproducibility was high both within and between laboratories, except for Hg and V. Exhaustive extraction and total digestion were generally comparable, between 87-111% for As, Cd, Co, and Pb. Total digestion exhibited lower variability than exhaustive extraction. Two types of microwave systems produced comparable results for most elements except Hg and Pb. The summation approach was comparable to direct analysis of tablets except for Hg and Cd, but summation demonstrated greater variability. XRF showed good agreement with ICP-MS and low replicate variability within labs. While the results were generally favorable, they demonstrate that some technical challenges remain to be addressed related to standardizing laboratory practices including interference correction strategies and selection of preparation methods. We discuss implications for method transfer between laboratories.

Journal of Trace Elements and MineralsVol. 12
RTI International (US), Rigaku (Japan) (JP), Thermo Fisher Scientific (United States) (US), Procter & Gamble (United States) (US), Perrigo (United States) (US)
Industry, innovation and infrastructure
Openalex Percentile: Top 96%
Extraction and Separation Processes
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.