Verification of the Analytical Performance of the Serum Lipoprotein (a) Assay on the Abbott Alinity ci®: Experience of the Biochemistry Laboratory at Mohammed VI University Hospital of Oujda

The verification of analytical methods, as required by the ISO 15189 standard, allows for the evaluation and comparison of a method’s performance against predefined analytical objectives. This study focuses on verifying the assay of Lipoprotein (a) on the Abbott® Alinity analyzer, contributing to accreditation and quality improvement in the laboratory. The assay was performed using immunoturbidimetry. The evaluation of analytical performance (repeatability and intermediate precision) was conducted using two quality control levels, with results analyzed via BYG Informatique software. The obtained results show satisfactory repeatability with CV1 = 1.16% and CV2 = 0.44%. Intra-laboratory reproducibility is also satisfactory, with CV1 = 2.16% and CV2 = 1.29%, in line with the manufacturer's specifications.

Authors

Publication Details

Journal
GSC Biological and Pharmaceutical Sciences
Published
2025-03-09
DOI
https://doi.org/10.30574/gscbps.2025.30.3.0092
Primary Topic
Lipoproteins and Cardiovascular Health
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Verification of the Analytical Performance of the Serum Lipoprotein (a) Assay on the Abbott Alinity ci®: Experience of the Biochemistry Laboratory at Mohammed VI University Hospital of Oujda

Dounia El Moujtahide, El-houcine Sebbar, Karim Jamal, Mohammed Choukri et al.
GSC Biological and Pharmaceutical Sciences
Lipoproteins and Cardiovascular Health
article

Verification of the Analytical Performance of the Serum Lipoprotein (a) Assay on the Abbott Alinity ci®: Experience of the Biochemistry Laboratory at Mohammed VI University Hospital of Oujda

Dounia El Moujtahide, El-houcine Sebbar, Karim Jamal, Mohammed Choukri, Oumaima kharkhach, Hasna Lekfif
article en

Abstract

The verification of analytical methods, as required by the ISO 15189 standard, allows for the evaluation and comparison of a method’s performance against predefined analytical objectives. This study focuses on verifying the assay of Lipoprotein (a) on the Abbott® Alinity analyzer, contributing to accreditation and quality improvement in the laboratory. The assay was performed using immunoturbidimetry. The evaluation of analytical performance (repeatability and intermediate precision) was conducted using two quality control levels, with results analyzed via BYG Informatique software. The obtained results show satisfactory repeatability with CV1 = 1.16% and CV2 = 0.44%. Intra-laboratory reproducibility is also satisfactory, with CV1 = 2.16% and CV2 = 1.29%, in line with the manufacturer's specifications.

GSC Biological and Pharmaceutical SciencesVol. 30(3)
Openalex Percentile: Top 93%
Lipoproteins and Cardiovascular Health
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.