A wireless intraoperative joint force sensing system for total hip arthroplasty

A wireless sensor-instrumented femoral head was developed that enables the intraoperative measurement of three force components in a total hip arthroplasty (THA). The design of the head was 32 mm in diameter with a 12/14 taper at the bottom, being compatible with commonly available THA. Inside the femoral head, there are three Hall effect sensors paired with permanent magnets, along with a microprocessor, a Bluetooth wireless module, a noncontact reed switch, and three lithium button batteries, all components were hermetically encapsulated. These components work together to measure and transmit the hip forces applied to the head during surgery. Through bench-top calibration, the system demonstrated a mean measurement error of approximately 8 % for forces up to 800 N. Additionally the practical applications of the prosthesis were tested on three hips from two cadavers undergoing THA. The hip forces were successively measured as the hip was passively moved, revealing insights into the potential for anterior or posterior dislocations with dislocation-prone positions. Understanding these intraoperative hip forces could aid surgeons in making the optimal decisions regarding implant placement and component selection, to minimize the risk of postoperative complications such as dislocation. Overall, this sensor-instrumented prosthesis will be useful for quantifying intraoperative hip forces.

Authors

Institutions

Publication Details

Journal
Medical Engineering & Physics
Published
2025-03-08
DOI
https://doi.org/10.1016/j.medengphy.2025.104325
Citations
1
Primary Topic
Orthopaedic implants and arthroplasty
Type
article
Field-Weighted Citation Impact
2.01

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

A wireless intraoperative joint force sensing system for total hip arthroplasty

Hiromasa Tanino, Ryo Mitsutake, Masaru Higa, Yusuke Hirai et al.
1 citations
Medical Engineering & Physics
Orthopaedic implants and arthroplasty
2.01
article

A wireless intraoperative joint force sensing system for total hip arthroplasty

Hiromasa Tanino, Ryo Mitsutake, Masaru Higa, Yusuke Hirai, Seita Inoue, Hiroshi Ito
article en
1 citations

Abstract

A wireless sensor-instrumented femoral head was developed that enables the intraoperative measurement of three force components in a total hip arthroplasty (THA). The design of the head was 32 mm in diameter with a 12/14 taper at the bottom, being compatible with commonly available THA. Inside the femoral head, there are three Hall effect sensors paired with permanent magnets, along with a microprocessor, a Bluetooth wireless module, a noncontact reed switch, and three lithium button batteries, all components were hermetically encapsulated. These components work together to measure and transmit the hip forces applied to the head during surgery. Through bench-top calibration, the system demonstrated a mean measurement error of approximately 8 % for forces up to 800 N. Additionally the practical applications of the prosthesis were tested on three hips from two cadavers undergoing THA. The hip forces were successively measured as the hip was passively moved, revealing insights into the potential for anterior or posterior dislocations with dislocation-prone positions. Understanding these intraoperative hip forces could aid surgeons in making the optimal decisions regarding implant placement and component selection, to minimize the risk of postoperative complications such as dislocation. Overall, this sensor-instrumented prosthesis will be useful for quantifying intraoperative hip forces.

Medical Engineering & PhysicsVol. 138(1)
Asahikawa Medical University (JP), University of Hyogo (JP), Hyogo University (JP)
Japan Society for the Promotion of Science
Openalex Percentile: Top 18%
Orthopaedic implants and arthroplasty
2.01
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

A wireless intraoperative joint force sensing system for total hip arthroplasty — Hiromasa Tanino, Ryo Mitsutake, et al. · Medical Engineering & Physics (2025) | TGRS Research Map | TGRS