A new pitfall in potassium testing: A case series of false elevation of hemolysis index in the presence of monoclonal protein

Hemolysis is the most common preanalytical interfering factor and a major source of error in clinical biochemistry analysis. Modern laboratories routinely assess hemolysis in conjunction with a laboratory information system algorithm that automatically adds a comment or rejects test values when appropriate. The hemolysis index has largely replaced visual inspection for assessing specimen quality prior to clinical chemistry testing. However, visual inspection of samples after centrifugation can provide valuable information about preanalytical interferences. In this report, we present a case series of inaccurate assessment of hemolysis index associated with the presence of monoclonal protein leading to inappropriate suppression of potassium results by the laboratory information system algorithm. These samples were originally identified as strongly hemolysed by the hemolysis index on the Atellica CH930 automated chemistry analyser. However, visual inspection revealed clear plasma, suggesting that the elevated hemolysis index was not due to hemolysis. This report highlights the need to be aware of potential monoclonal protein interference with the hemolysis index for better interpretation of results, and proposes three warning signs to detect this type of anomaly.

Authors

Institutions

Publication Details

Journal
Clinica Chimica Acta
Published
2025-03-05
DOI
https://doi.org/10.1016/j.cca.2025.120224
Primary Topic
Clinical Laboratory Practices and Quality Control
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

A new pitfall in potassium testing: A case series of false elevation of hemolysis index in the presence of monoclonal protein

Hélène Kerspern, Jean‐Christophe Ianotto, Clément Capaldo, Cyril Leven et al.
Clinica Chimica Acta
Clinical Laboratory Practices and Quality Control
article

A new pitfall in potassium testing: A case series of false elevation of hemolysis index in the presence of monoclonal protein

Hélène Kerspern, Jean‐Christophe Ianotto, Clément Capaldo, Cyril Leven, Jean-Richard Eveillard
article en

Abstract

Hemolysis is the most common preanalytical interfering factor and a major source of error in clinical biochemistry analysis. Modern laboratories routinely assess hemolysis in conjunction with a laboratory information system algorithm that automatically adds a comment or rejects test values when appropriate. The hemolysis index has largely replaced visual inspection for assessing specimen quality prior to clinical chemistry testing. However, visual inspection of samples after centrifugation can provide valuable information about preanalytical interferences. In this report, we present a case series of inaccurate assessment of hemolysis index associated with the presence of monoclonal protein leading to inappropriate suppression of potassium results by the laboratory information system algorithm. These samples were originally identified as strongly hemolysed by the hemolysis index on the Atellica CH930 automated chemistry analyser. However, visual inspection revealed clear plasma, suggesting that the elevated hemolysis index was not due to hemolysis. This report highlights the need to be aware of potential monoclonal protein interference with the hemolysis index for better interpretation of results, and proposes three warning signs to detect this type of anomaly.

Clinica Chimica ActaVol. 571
Inserm (FR), Université de Bretagne Occidentale (FR), Centre Hospitalier Régional Universitaire de Brest (FR)
Openalex Percentile: Top 94%
Clinical Laboratory Practices and Quality Control
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

A new pitfall in potassium testing: A case series of false elevation of hemolysis index in the presence of monoclonal protein — Hélène Kerspern, Jean‐Christophe Ianotto, et al. · Clinica Chimica Acta (2025) | TGRS Research Map | TGRS